Holding Registers (4xxxx Addresses)
34
Address Register Function
Size
Min.
Value
Max.
Value
Default
Value Details
40022
Cleaning post wash time
uint16_t
5
999
20
[seconds]
40023
Cleaning Recovery time
uint16_t
5
120
10
[seconds]
40024
Cleaning Wash Cycles
uint8_t
1
10
1
40025
Cleaning Rinse only cycles
uint8_t
0
10
0
40026
Cleaning Schedule 1
st
time to start hour uint8_t
0
23
0
40027
Cleaning Schedule 1
st
time to start
minute
uint8_t
0
59
0
40028
Cleaning Schedule 1
st
time to start
enabled
uint8_t
0
1
0
0
= start disabled
1
= start enabled
40029
Cleaning Schedule 2
nd
time to start hour uint8_t
0
23
0
40030
Cleaning Schedule 2
nd
time to start
minute
uint8_t
0
59
0
40031
Cleaning Schedule 2
nd
time to start
enabled
uint8_t
0
1
0
0
= start disabled
1
= start enabled
40032
Cleaning Schedule 3
rd
time to start hour uint8_t
0
23
0
40033
Cleaning Schedule 3
rd
time to start
minute
uint8_t
0
59
0
40034
Cleaning Schedule 3
rd
time to start
enabled
uint8_t
0
1
0
0
= start disabled
1
= start enabled
40035
Input1 Function
uint8_t
0
2
0
0
= Disabled
1
= Hold
2
= Clean
40036
Input1 Active Level
uint8_t
0
1
0
0
= Low
1
= High
40037
Input2 Function
uint8_t
0
2
0
0
= Disabled
1
= Hold
2
= Clean
40038
Input2 Active Level
uint8_t
0
1
0
0
= Low
1
= High
40041
Relay 1 function
uint8_t
0
5
0
0
= Disabled
1
= CtrlSetP1
2
= CtrlSetP2
3
= Rinse
4
= Wash
5
= Hold
40042
Relay 2 function
uint8_t
0
5
0
40043
Relay 3 function
uint8_t
0
5
0
40044
Relay 4 function
uint8_t
0
5
0
40045
Relay 5 function
uint8_t
0
5
0
40046
Hold delay
uint8_t
0
99
5
[seconds]
40047
AnOut1 mode
uint8_t
0
1
0
0
= Disabled
1
= Track Channel
40048
AnOut1 data Channel
uint8_t
0
1
0
0
= CH1 (Probe 1)
1
= CH2 (Probe 2)
40049
AnOut1 parameter to follow
uint8_t
0
4
0
0
= Ctrl SetP1
1
= Ctrl SetP2
2
= Main parameter
3
= Temper. Param.
4
= Auxiliary Param.
40050
AnOut1 mA range
uint8_t
0
1
0
0
= “0‑20 mA”
1
= “4‑20 mA”
40051
AnOut1 Value in Hold option
uint8_t
0
1
0
0
= “Last value”
1
= “Fix value”