The flashing digits indicate that:

1) the month can be set
2) the day can be set

“Storage”: this prompt appears for 1 second at the end of the

calibration procedure

, to indicate that the calibration data has

been stored.

“Error”: the concentration of the calibration solution used is not
correct. Repeat the 

calibration procedure

 with the right standard

solution, and verify it is not expired. If the calibration procedure fails
again, contact your dealer or the nearest Hanna Customer Service
Center.

ERROR  MESSAGES

a) on zero reading

This indicates that the zeroing procedure failed due to a low signal-to-
noise ratio. In this case press ZERO again.

“No Light”: the instrument cannot adjust the light level. Please check
that the sample does not contain any debris.

“Light Low”: there is not enough light to perform a measurement.
Please check the preparation of the zero cuvet.

“Light High”: there is too much light to perform a measurement.
Please check the preparation of the zero cuvet.

b) on sample reading

There is too much light for the sample measurement. Please check if
the right sample cuvet is inserted.

“Inverted”: the sample and the zero cuvet are inverted.

“Zero”: a zero reading was not taken. Follow the instruction in the

measurement procedure

 for zeroing the meter.

11

Under range. A blinking "0.000" indicates that the sample absorbs
less light than the zero reference. Check the procedure and make
sure you use the same cuvet for reference (zero) and measurement.

A  flashing value of the maximum concentration indicates an
over range condition. The concentration of the sample is beyond
the programmed range: dilute the sample and re-run the test.

A flashing value lower than the maximum concentration indicates
a  low signal-to-noise ratio condition. In this case accuracy of the
result is not guaranteed. Repeat the 

measurement procedure

.

The instructions listed below should be carefully followed during testing to ensure best accuracy.

• For a correct filling of the cuvet: the liquid in the

cuvet forms a convexity on the top; the bottom
of this convexity must be at the same level of

the 10 mL mark.

• Proper use of the powder reagent packet:

(a) use scissors to open the powder packet;
(b) push the edges of the packet to form a spout;
(c) pour out the content of the packet.

(a)

(b)

(c)

GENERAL TIPS FOR AN ACCURATE MEASUREMENT

Summary of Contents for HI 95762C

Page 1: ...tion Manual HI 95762C Free Chlorine Meter for Drinking Water Application w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives 20 MAN95762R2 05 06 w w w h a n n a i n s...

Page 2: ...ake all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid damages or burns do not put the ins...

Page 3: ...ESSORIES HI 95762 11 CAL CHECK Standard Cuvets 1 set HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 721310 9V battery 10 pcs HI 731318 Tissue for wiping cuvets 4 pcs HI 731331 Glass cuvets...

Page 4: ...62 portable meter measures the free chlorine Cl2 content in water samples in the 0 000 to 0 500 mg L ppm range The method is an adaption of the Standard Method 4500 Cl G for drinking water The reagent...

Page 5: ...L chlorine and a representative lot of reagent an operator obtained with a single instrument a standard deviation of 0 004 mg L SPECIFICATIONS PRECISION AND ACCURACY 16 CALIBRATION PROCEDURE Turn the...

Page 6: ...band interference filters to guarantee both high performance and reliable results HI 95 series block diagram optical layout PRINCIPLE OF OPERATION 15 Turn the meter on by pressing ON OFF When the LCD...

Page 7: ...vet walls with hands Furthermore in order to maintain the same conditions during the zeroing and the measuring phases it is necessary to close the cuvet to prevent any contamination 1 Liquid Crystal D...

Page 8: ...3 the Month and Day icons appear above the number of month and day when a date is displayed 4 the blinking Battery icon appears when the battery voltage is getting low 13 MEASUREMENT PROCEDURE Turn t...

Page 9: ...ibration is selected GUIDE TO DISPLAY CODES 12 It is important that the sample does not contain any debris This would corrupt the reading Each time the cuvet is used the cap must be tightened to the s...

Page 10: ...the right sample cuvet is inserted Inverted the sample and the zero cuvet are inverted Zero a zero reading was not taken Follow the instruction in the measurement procedure for zeroing the meter 11 U...

Reviews: