background image

9

This prompt appears for 1 second each time the instrument is
turned on.

This indicates that the meter is in a ready state and zeroing can be
performed.

Sampling in Progress. This flashing prompt appears each time the
meter is performing a measurement.

This indicates that the meter is in a zeroed state and measurement
can be performed.

Light over range: the cuvet is not inserted correctly and an excess

ambient light is reaching the detector. If the cover is properly
installed, then contact your dealer or the nearest Hanna Customer
Service Center.

The blinking "BAT" indicates that the battery voltage is getting low
and the battery needs to be replaced.

This indicates that the battery is dead and must be replaced. Once
this indication is displayed, the meter will lock up. Change the

battery and restart.

“Configuration”: this indicates that the meter has lost its
configuration. Contact your dealer or the nearest Hanna Customer
Service Center.

CALIBRATION MODE  MESSAGES

The date of the last calibration appears on the LCD each time the

calibration mode

 is entered. If calibration is performed for the first

time, “F.CAL” appears.

This indicates that the factory calibration is selected.

GUIDE TO DISPLAY CODES

12

• It is important that the sample does not contain

any debris. This would corrupt the reading.

• Each time the cuvet is used, the cap must be

tightened to the same degree.

• Whenever the cuvet is placed into the measurement

cell, it must be dry outside, and completely free of
fingerprints, oil or dirt. Wipe it thoroughly with

HI 731318

 (tissue for wiping cuvets, see chapter

ACCESSORIES) or a lint-free cloth prior to insertion.

• Shaking the cuvet can generate bubbles in the

sample, causing higher readings. To obtain accurate
measurements, remove such bubbles by swirling
or by gently tapping the vial.

• Do not let the reacted sample stand too long after

reagent is added, or accuracy will be lost.

• It is possible to take multiple readings in a row,

but it is recommended to take a new zero reading
for each sample and to use the same cuvet for
zeroing and measurement.

• After the reading it is important to discard

immediately the sample, otherwise the glass might

become permanently stained.

• All the reaction times reported in this manual are

referred to 20°C (68°F). As a general rule of
thumb, they should be doubled at 10°C (50°F)
and halved at 30°C (86°F).

• In order to maximize accuracy, prior to a

measurement follow the 

validation procedure

 to

be sure that the instrument is properly calibrated.

If necessary, calibrate the instrument.

Summary of Contents for HI 95762C

Page 1: ...tion Manual HI 95762C Free Chlorine Meter for Drinking Water Application w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives 20 MAN95762R2 05 06 w w w h a n n a i n s...

Page 2: ...ake all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid damages or burns do not put the ins...

Page 3: ...ESSORIES HI 95762 11 CAL CHECK Standard Cuvets 1 set HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 721310 9V battery 10 pcs HI 731318 Tissue for wiping cuvets 4 pcs HI 731331 Glass cuvets...

Page 4: ...62 portable meter measures the free chlorine Cl2 content in water samples in the 0 000 to 0 500 mg L ppm range The method is an adaption of the Standard Method 4500 Cl G for drinking water The reagent...

Page 5: ...L chlorine and a representative lot of reagent an operator obtained with a single instrument a standard deviation of 0 004 mg L SPECIFICATIONS PRECISION AND ACCURACY 16 CALIBRATION PROCEDURE Turn the...

Page 6: ...band interference filters to guarantee both high performance and reliable results HI 95 series block diagram optical layout PRINCIPLE OF OPERATION 15 Turn the meter on by pressing ON OFF When the LCD...

Page 7: ...vet walls with hands Furthermore in order to maintain the same conditions during the zeroing and the measuring phases it is necessary to close the cuvet to prevent any contamination 1 Liquid Crystal D...

Page 8: ...3 the Month and Day icons appear above the number of month and day when a date is displayed 4 the blinking Battery icon appears when the battery voltage is getting low 13 MEASUREMENT PROCEDURE Turn t...

Page 9: ...ibration is selected GUIDE TO DISPLAY CODES 12 It is important that the sample does not contain any debris This would corrupt the reading Each time the cuvet is used the cap must be tightened to the s...

Page 10: ...the right sample cuvet is inserted Inverted the sample and the zero cuvet are inverted Zero a zero reading was not taken Follow the instruction in the measurement procedure for zeroing the meter 11 U...

Reviews: