9
This prompt appears for 1 second each time the instrument is turned on.
The parameter code “724” indicates that the meter is in a ready state
and zeroing can be performed.
Sampling in Progress. This
flashing prompt
appears each time the meter
is performing a measurement.
This indicates that the meter is in a zeroed state and measurement can
be performed.
The blinking "BAT" indicates that the battery voltage is getting low and
the battery needs to be replaced.
This indicates that the battery is dead and must be replaced. Once this
indication is displayed, the meter will lock up. Change the battery and
restart.
“Configuration”: this indicates that the meter has lost its configuration.
Contact your dealer or the nearest Hanna Customer Service Center.
CALIBRATION MODE MESSAGES
The date of the last calibration appears on the LCD each time the
calibration mode
is entered. If calibration is performed for the first time,
“F.CAL” appears.
This indicates that the factory calibration is selected.
GUIDE TO DISPLAY CODES
2
2
2
2
2
1
1
1
1
1
The flashing digits indicate that:
1) the month can be set
2) the day can be set
“Storage”: this prompt appears for 1 second at the end of the calibration
procedure, to indicate that the calibration data has been stored.
16
or
Note:
at this point it is possible to reset the
instrument to restore FACTORY
CALIBRATION, see instructions page 17.
• Place the CAL CHECK™ Standard Cuvet
A
into the holder and ensure that the notch on
the cap is positioned securely into the groove.
• Press ZERO and “SIP” will blink on the
display.
• After a few seconds the display will
show “-0.0-”. The meter is now zeroed and
ready for calibration.
• Remove the cuvet and place the CAL CHECK™
Standard HI 95724-11 Cuvet
B
into the
holder. Ensure that the notch on the cap is
positioned securely into the groove.
• Press READ and “SIP” will blink during
measurement.
• The instrument will show for three seconds
the CAL CHECK™ standard value.
Then the date of last calibration (e.g.:
“01.08”) appears on the display, or “01.01”
if the factory calibration was selected. In
both cases the number of the month is
blinking, ready for date input.
Note:
if the display shows “ERR” the calibration procedure failed. Verify that both CAL CHECK™
Standards HI 95724-11 Cuvets, A and B, are free from fingerprints or dirt and that they
are inserted correctly.