Hanna Instruments HI 95724C Instruction Manual Download Page 7

7

FUNCTIONAL DESCRIPTION

1) Liquid Crystal Display

2) Cuvet Holder
3) Cuvet alignment indicator
4) ON/OFF key

5) ZERO key

6) READ key
7) CAL CHECK key

A microprocessor controlled special tungsten lamp emits radiation which is first optically
conditioned and beamed to the sample contained in the cuvet. The optical path is fixed by the
diameter of the cuvet. Then the light is spectrally filtered to a narrow spectral bandwidth, to
obtain a light beam of intensity 

I

o

 or 

I

.

The photoelectric cell collects the radiation 

I

 that is not absorbed by the sample and converts

it into an electric current, producing a potential in the mV range.
The microprocessor uses this potential to convert the incoming value into the desired measuring
unit and to display it on the LCD.
The measurement process is carried out in two phases: first the meter is zeroed and then the
actual measurement is performed.
The cuvet has a very important role because it is an optical element and thus requires particular
attention. It is important that both the measurement and the calibration (zeroing) cuvets are
optically identical to provide the same measurement conditions. Whenever possible use the same
cuvet for both. It is necessary that the surface of the cuvet is clean and not scratched. This to
avoid measurement interference due to unwanted reflection and absorption of light. It is
recommended not to touch the cuvet walls with hands.
Furthermore, in order to maintain the same conditions during the zeroing and the measuring
phases, it is necessary to close the cuvet to prevent any contamination.

18

Battery replacement must only take place in a
non-hazardous area.
Simply slide off the battery cover on the back of
the meter. Detach the battery from the terminals
and attach a fresh 9V battery while paying
attention to the correct polarity. Insert the battery
and replace the cover.

9V

SLIDE

OFF

BATTERY REPLACEMENT

Summary of Contents for HI 95724C

Page 1: ...1 w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives Instruction Manual HI 95724C Free Total Chlorine ISM 24 MAN95724C 01 04 w w w h a n n a i n s t c o m...

Page 2: ...ER NOTES 23 Dear Customer Thank you for choosing a Hanna product This manual will provide you with the necessary information for the correct use of the instrument Please read it carefully before using...

Page 3: ...ll packing material until you are sure that the instrument works correctly Any defective item must be returned in its original packing 22 Hanna publishes a wide range of catalogs and handbooks for an...

Page 4: ...aced into the measurement cell It is designed to fit a cuvet with a larger neck making it easier to add both sample and reagents The cuvet is made from special optical glass to obtain best results The...

Page 5: ...The figure explains these definitions In a laboratory using a standard solution of 1 00 mg L chlorine and a representative lot of reagents an operator obtained with a single instrument a standard devi...

Page 6: ...s based on the possibility to develop an absorbing compound from a specific chemical reaction between sample and reagents Given that the absorption of a compound strictly depends on the wavelength of...

Page 7: ...then the actual measurement is performed The cuvet has a very important role because it is an optical element and thus requires particular attention It is important that both the measurement and the...

Page 8: ...ppears when the battery voltage is getting low 17 FACTORY CALIBRATION RESET It is possible to restore factory calibration DATE INPUT Keep READ pressed to scroll to the desired month number 01 12 When...

Page 9: ...icate that 1 the month can be set 2 the day can be set Storage this prompt appears for 1 second at the end of the calibration procedure to indicate that the calibration data has been stored 16 or Note...

Page 10: ...e sample is beyond the programmed range dilute the sample and re run the test A flashing value lower than the maximum concentration indicates a low signal to noise ratio condition In this case accurac...

Page 11: ...318 tissue for wiping cuvets see chapter Accessories or a lint free cloth prior to insertion a b 14 VALIDATION PROCEDURE Use the validation procedure to ensure that the instrument is properly calibrat...

Page 12: ...that the notch on the cap is positioned securely into the groove Press ZERO and SIP will blink on the display After a few seconds the display will show 0 0 The meter is now zeroed and ready for measu...

Reviews: