Hanna Instruments HI 95724C Instruction Manual Download Page 11

11

GENERAL TIPS FOR AN ACCURATE MEASUREMENT

The instructions listed below should be carefully followed during testing
to ensure best accuracy.

• For a correct filling of the cuvet: the liquid in the cuvet

forms a convexity on the top; the bottom of this convex-
ity must be at the same level of the 10 mL mark.

• Proper use of the dropper:

(a) to get good reproducible results, tap the dropper on the table for

several times and wipe  the outside of the dropper tip with a
cloth.

(b) always keep the dropper bottle in a vertical position while

dosing the reagent.

• It is important that the sample does not contain any debris. This

would corrupt the reading.

• In order to avoid reagent leaking and to obtain more

accurate measurements, it s recommended to close the
cuvet first with the supplied HDPE plastic stopper and
then with the black cap.

• Each time the cuvet is used, the cap must be tightened to the same

degree.

• Whenever the cuvet is placed into the measurement

cell, it must be dry outside, and completely free of
fingerprints, oil or dirt. Wipe it thoroughly with

HI 731318

 (tissue for wiping cuvets, see chapter

“Accessories”) or a lint-free cloth prior to insertion.

(

a

)

(

b

)

14

VALIDATION PROCEDURE

Use the validation procedure to ensure that the instrument is properly calibrated.

Warning:

 do not validate or calibrate the instrument with standard solutions other than the Hanna

CAL CHECK™ Standards, otherwise erroneous results will be obtained. For accurate Validation and
Calibration, please perform test at room temperature: 18 to 25°C (64.5 to 77.0°F).

• Turn the meter on by pressing ON/OFF.

• When the LCD displays “724”, it is ready.

• Place the CAL CHECK™ Standard Cuvet 

A

into the holder and ensure that the notch on
the cap is positioned securely into the groove.

• Press ZERO and “SIP” will blink on the

display.

• After a few seconds the display will

show “-0.0-”. The meter is now zeroed and
ready for validation.

INTERFERENCES

• Positive error: - Bromine

- Chlorine dioxide
- Iodine
- Ozone
- Oxidized Manganese and Chromium

• Alkalinity above 250 mg/L CaCO

or acidity above 150 mg/L CaCO

3

 will not reliably develop

the full amount of color or it may rapidly fade.
To resolve this, neutralize the sample with diluted HCl or NaOH.

• In case of water with hardness greater than 500 mg/L CaCO

3

, shake the sample for

approximately 2 minutes after adding the powder reagent.

Summary of Contents for HI 95724C

Page 1: ...1 w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives Instruction Manual HI 95724C Free Total Chlorine ISM 24 MAN95724C 01 04 w w w h a n n a i n s t c o m...

Page 2: ...ER NOTES 23 Dear Customer Thank you for choosing a Hanna product This manual will provide you with the necessary information for the correct use of the instrument Please read it carefully before using...

Page 3: ...ll packing material until you are sure that the instrument works correctly Any defective item must be returned in its original packing 22 Hanna publishes a wide range of catalogs and handbooks for an...

Page 4: ...aced into the measurement cell It is designed to fit a cuvet with a larger neck making it easier to add both sample and reagents The cuvet is made from special optical glass to obtain best results The...

Page 5: ...The figure explains these definitions In a laboratory using a standard solution of 1 00 mg L chlorine and a representative lot of reagents an operator obtained with a single instrument a standard devi...

Page 6: ...s based on the possibility to develop an absorbing compound from a specific chemical reaction between sample and reagents Given that the absorption of a compound strictly depends on the wavelength of...

Page 7: ...then the actual measurement is performed The cuvet has a very important role because it is an optical element and thus requires particular attention It is important that both the measurement and the...

Page 8: ...ppears when the battery voltage is getting low 17 FACTORY CALIBRATION RESET It is possible to restore factory calibration DATE INPUT Keep READ pressed to scroll to the desired month number 01 12 When...

Page 9: ...icate that 1 the month can be set 2 the day can be set Storage this prompt appears for 1 second at the end of the calibration procedure to indicate that the calibration data has been stored 16 or Note...

Page 10: ...e sample is beyond the programmed range dilute the sample and re run the test A flashing value lower than the maximum concentration indicates a low signal to noise ratio condition In this case accurac...

Page 11: ...318 tissue for wiping cuvets see chapter Accessories or a lint free cloth prior to insertion a b 14 VALIDATION PROCEDURE Use the validation procedure to ensure that the instrument is properly calibrat...

Page 12: ...that the notch on the cap is positioned securely into the groove Press ZERO and SIP will blink on the display After a few seconds the display will show 0 0 The meter is now zeroed and ready for measu...

Reviews: