Hanna Instruments HI 95724C Instruction Manual Download Page 10

10

“Error”: the concentration of the calibration solution used is not correct.
Repeat the 

calibration procedure

 with the right standard solution, and

verify it is not expired. If the calibration procedure fails again, contact your
dealer or the nearest Hanna Customer Service Center.

ERROR MESSAGES

a) on zero reading

This indicates that the zeroing procedure failed due to a low signal-to-noise
ratio. In this case press ZERO again.

“No Light”: the instrument cannot adjust the light level. Please check that
the sample does not contain any debris.

“Light Low”: there is not enough light to perform a measurement. Please
check the preparation of the zero cuvet.

“Light High”: there is too much light to perform a measurement. Please check
the preparation of the zero cuvet.

b) on sample reading

There is too much light for the sample measurement. Please check if the right
sample cuvet is inserted.

“Inverted”: the sample and the zero cuvet are inverted.

“Zero”: a zero reading was not taken. Follow the instruction in the

measurement procedure

 for zeroing the meter.

Under range. A blinking "0.00" indicates that the sample absorbs less
light than the zero reference. Check the procedure and make sure you
use the same cuvet for reference (zero) and measurement.

A  flashing value of the maximum concentration indicates an over range
condition. The concentration of the sample is beyond the programmed
range: dilute the sample and re-run the test.

A flashing value lower than the maximum concentration indicates a  low
signal-to-noise ratio condition. In this case accuracy of the result is not
guaranteed. Repeat the 

measurement procedure

.

15

• Remove the cuvet and place the CAL

CHECK™ Standard HI 95724-11 Cuvet 

B

into the holder. Ensure that the notch on
the cap is positioned securely into the
groove.

• Press CAL CHECK and “SIP” will blink during

measurement.

• Wait for a few seconds and the display will

show the validation standard value.
The reading should be within specifications
as reported on the CAL CHECK™ Standard
Certificate. If the value found is out of
specifications, please check that the cuvets
are free of fingerprints, oil or dirt and
repeat validation. If results are still found
out of specifications then recalibrate the
instrument.

CALIBRATION PROCEDURE

Note

: It is possible to interrupt calibration procedure at any time by pressing ON/OFF.

Warning:

 do not validate or calibrate the instrument with standard solutions other than the Hanna

CAL CHECK™ Standards, otherwise erroneous results will be obtained. For accurate Validation and
Calibration, please perform test at room temperature: 18 to 25°C (64.5 to 77.0°F).

• Turn the meter on by pressing ON/OFF.

• When the LCD displays “724”, it is ready.

• Enter the 

calibration mode

 by holding CAL

CHECK 

for three seconds

.

• The date of the last calibration appears (e.g.:

month “01”, day “08”). “F.CAL” means that
the factory calibration is selected.

or

Summary of Contents for HI 95724C

Page 1: ...1 w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives Instruction Manual HI 95724C Free Total Chlorine ISM 24 MAN95724C 01 04 w w w h a n n a i n s t c o m...

Page 2: ...ER NOTES 23 Dear Customer Thank you for choosing a Hanna product This manual will provide you with the necessary information for the correct use of the instrument Please read it carefully before using...

Page 3: ...ll packing material until you are sure that the instrument works correctly Any defective item must be returned in its original packing 22 Hanna publishes a wide range of catalogs and handbooks for an...

Page 4: ...aced into the measurement cell It is designed to fit a cuvet with a larger neck making it easier to add both sample and reagents The cuvet is made from special optical glass to obtain best results The...

Page 5: ...The figure explains these definitions In a laboratory using a standard solution of 1 00 mg L chlorine and a representative lot of reagents an operator obtained with a single instrument a standard devi...

Page 6: ...s based on the possibility to develop an absorbing compound from a specific chemical reaction between sample and reagents Given that the absorption of a compound strictly depends on the wavelength of...

Page 7: ...then the actual measurement is performed The cuvet has a very important role because it is an optical element and thus requires particular attention It is important that both the measurement and the...

Page 8: ...ppears when the battery voltage is getting low 17 FACTORY CALIBRATION RESET It is possible to restore factory calibration DATE INPUT Keep READ pressed to scroll to the desired month number 01 12 When...

Page 9: ...icate that 1 the month can be set 2 the day can be set Storage this prompt appears for 1 second at the end of the calibration procedure to indicate that the calibration data has been stored 16 or Note...

Page 10: ...e sample is beyond the programmed range dilute the sample and re run the test A flashing value lower than the maximum concentration indicates a low signal to noise ratio condition In this case accurac...

Page 11: ...318 tissue for wiping cuvets see chapter Accessories or a lint free cloth prior to insertion a b 14 VALIDATION PROCEDURE Use the validation procedure to ensure that the instrument is properly calibrat...

Page 12: ...that the notch on the cap is positioned securely into the groove Press ZERO and SIP will blink on the display After a few seconds the display will show 0 0 The meter is now zeroed and ready for measu...

Reviews: