MEASUREMENT PROCEDURE

• Turn the meter on by pressing ON/OFF.

• When the LCD displays “724”, it is ready.

• Do not let the reacted sample stand too long after reagent is added, or accuracy will be lost.

• It is possible to take multiple readings in a row, but it is recommended to take a new zero

reading for each sample and to use the same cuvet for zeroing and measurement.

• After the reading it is important to discard immediately the sample, otherwise the glass

might become permanently stained.

• All the reaction times reported in this manual are referred to 20°C (68°F). As a general rule

of thumb, they should be doubled at 10°C (50°F) and halved at 30°C (86°F).

• In order to maximize accuracy, prior to a measurement follow the validation procedure to be

sure that the instrument is properly calibrated. If necessary, calibrate the instrument.

• Fill the cuvet with 10 mL of unreacted

sample, up to the mark, and replace the
cap.

• Place the cuvet into the holder and ensure

that the notch on the cap is positioned
securely into the groove.

• Press ZERO and “SIP” will blink on the

display.

• After a few seconds the display will show

“-0.0-”. The meter is now zeroed and
ready for measurement.

• Remove the cuvet.

13

For free chlorine measurement

: add 3 drops

of HI 93701A-T DPD1 indicator and 3
drops of HI 93701B-T DPD1 buffer to an

empty cuvet

.

Immediately

 add 10 mL of unreacted

sample. Replace the cap and shake gently.

Immediately 

reinsert the cuvet into the

holder and ensure that the notch on the
cap is positioned securely into the groove.

• Press READ and “SIP” will blink during

measurement.

• The instrument directly displays

concentration in mg/L of free chlorine on
the Liquid Crystal Display.

• Then, 

for total chlorine measurement

, remove

the cuvet and add to the reacted sample 3
drops of HI 93701C-T DPD3 solution.
Replace the cap and swirl gently.

• Reinsert the cuvet into the holder and

ensure that the notch on the cap is
positioned securely into the groove.

• Wait for 2’ and 30” then press READ and

“SIP” will blink during measurement.

• The instrument directly displays

concentration in mg/L of total chlorine on
the Liquid Crystal Display.

Summary of Contents for HI 95724C

Page 1: ...1 w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives Instruction Manual HI 95724C Free Total Chlorine ISM 24 MAN95724C 01 04 w w w h a n n a i n s t c o m...

Page 2: ...ER NOTES 23 Dear Customer Thank you for choosing a Hanna product This manual will provide you with the necessary information for the correct use of the instrument Please read it carefully before using...

Page 3: ...ll packing material until you are sure that the instrument works correctly Any defective item must be returned in its original packing 22 Hanna publishes a wide range of catalogs and handbooks for an...

Page 4: ...aced into the measurement cell It is designed to fit a cuvet with a larger neck making it easier to add both sample and reagents The cuvet is made from special optical glass to obtain best results The...

Page 5: ...The figure explains these definitions In a laboratory using a standard solution of 1 00 mg L chlorine and a representative lot of reagents an operator obtained with a single instrument a standard devi...

Page 6: ...s based on the possibility to develop an absorbing compound from a specific chemical reaction between sample and reagents Given that the absorption of a compound strictly depends on the wavelength of...

Page 7: ...then the actual measurement is performed The cuvet has a very important role because it is an optical element and thus requires particular attention It is important that both the measurement and the...

Page 8: ...ppears when the battery voltage is getting low 17 FACTORY CALIBRATION RESET It is possible to restore factory calibration DATE INPUT Keep READ pressed to scroll to the desired month number 01 12 When...

Page 9: ...icate that 1 the month can be set 2 the day can be set Storage this prompt appears for 1 second at the end of the calibration procedure to indicate that the calibration data has been stored 16 or Note...

Page 10: ...e sample is beyond the programmed range dilute the sample and re run the test A flashing value lower than the maximum concentration indicates a low signal to noise ratio condition In this case accurac...

Page 11: ...318 tissue for wiping cuvets see chapter Accessories or a lint free cloth prior to insertion a b 14 VALIDATION PROCEDURE Use the validation procedure to ensure that the instrument is properly calibrat...

Page 12: ...that the notch on the cap is positioned securely into the groove Press ZERO and SIP will blink on the display After a few seconds the display will show 0 0 The meter is now zeroed and ready for measu...

Reviews: