29
MEMORY ORGANIZATION
(HI 92810 & HI 92840 ONLY)
Capacity:
16,000 data samples, divided into
16 pages.
Data capacity per page:
1000 data samples with 1 chan-
nel monitored;
500 data samples with 2 chan-
nels monitored (four channels
versions only);
250 data samples with 3 or 4
channels monitored (four chan-
nels versions only).
Each time a new logging period starts, it
automatically starts from a new page.
If “LOGGING” is still on, and the available
page is “0” the meter will overwrite the first
LOT DATA in the existing meter memory.
During logging the meter automatically re-
turns to the oldest page in the memory and if
it contains data, it will overwrite. In this case
the first log will not correspond to the oldest
set of data
It is recommended to periodically “clean” the
memory. Save the data in a PC if you need
to keep a record and then disconnect the
batteries for about 1 minute. If you do this,
remember to set the date and time, once the
batteries have been connected again.
ATTENTION
Data are stored into memory until batteries
are removed.
If replacement of the batteries is needed and
data are not to be lost, plug the adapter in
and proceed with batteries replacement as
described below. Only once batteries have
been replaced it is possible to unplug the
adapter without loosing the stored data.