18
CH
°C
LOG
To reactivate the display
press the TEMP key.
The printout provides the following informa-
tion:
a - Date (DD-MM-YY)
b - A running log number
c - A running sample number
(in that particular log)
d - Printing interval indicator in
minutes
e - Time (HH-MM)
f - Temperature value(s)
It is always possible to switch from the log-
ging with printing function to the logging only
function. Press the ALT and the PAPER keys
at the same time and the "LOG" symbol will
start to blink to indicate that the data are now
stored only into memory.
Notes:
•
It is recommended to use the adapter
during logging with printing mode, espe-
cially when many printout are going to be
taken.
•
Before proceeding with logging with print-
ing, make sure there is enough paper for
your measurements. When the paper is
finished the meter will not advise the op-
erator and the printouts could be lost. If
this happens, data will continue to be
stored into memory, and it is always pos-
sible to print them at different time (see
page 22).
10-01-96
==07== 0005M
0014 *14.25
#1 25.0 °C
#2 27.3 °C
.
.
b
c
d
a
e
f
UP
PAPER
ALT
TEMP