16
CH
°C
LOG
To continue logging without printing, press
now the ALT and the PAPER keys at the
same time and the "LOG" symbol on display
will start to blink.
After approximately 5 minutes the display will
switch off but the logging function remains
active.
To reactivate the display
press the TEMP key.
If you wish to re-start printing press the ALT
and PAPER keys simultaneously again.
The printout provides the following informa-
tion:
a - Date (DD-MM-YY)
b - A running log number
c - Printing interval indicator in minutes
d - Time (HH-MM)
e - Temperature value(s)
f - Running sample number (in that particular
log)
Notes:
•
Once in the logging mode,
the interval cannot be
changed. Exit the logging
mode first (pressing the
ALT and the LOG keys) before setting the
new interval.
•
If the PRINT key is
pressed while in logging
mode, a printout is pro-
duced without affecting the
running sample number.
UP
PAPER
ALT
UP
DOWN
TIME
LOG
ALT
TEMP
TEMP
10-01-96
==07== 0005M
0001 *14.25
#1 25.0 °C
#2 27.3 °C
.
.
b
f
c
a
d
e