29

28

• Leaving the probe in the "RH 75.4%" chamber, adjust the low

humidity trimmer until the display shows 75.4%.

• The humidity calibration is now complete.
Spare saturation salts are also available: HI 7111/P (LiCl) for low
range humidity, and HI 7121/P (NaCl) for high range humidity.

DATA TRANSFER TO PC

DATA TRANSFER TO PC

DATA TRANSFER TO PC

DATA TRANSFER TO PC

DATA TRANSFER TO PC

(HI 91610 only)

HI 91610 

contains infrared emitting circuitry.

Set the meter to TIME mode and place it on a

HI 9200 

Infrared Transmitter (ensuring that

the infrared LEDs are placed on top of each
other). The logged data can be downloaded
to your PC through the 

HI 9200

's RS232 port.

During the data transfer the instrument dis-
plays "r 232".

Using the 

HI 9200 

Infrared Transmitter, all recorded data can be fed

to your Personal Computer for easy reproduction, storage or elabora-
tion without having to connect and disconnect cables between the
meter and your PC.
Data can be further elaborated with the new optional 

HI 92000

Windows

®

 compatible application software.

TIME

Windows

®

 is a registered Trademark of "Microsoft Co."

• Press the ON/OFF key (for HI 9161/

HI  9261) or the RANGE key (for
HI  91610) to switch the instrument on.

• Press the RANGE key to display the RH%

reading on the primary LCD.

• Turn the low humidity trimmer (see Functional Description) until a

value of 0.0% is reached (readings between 0.0% and 1.0% RH
are acceptable).

• Remove the probe and tightly seal the chamber containing the

LiCl solution.

• Remove the cap from the "RH75.4%"

chamber containing the NaCl solution
and insert the probe, ensuring that it
does not come in contact with the liquid.

• Wait for the measurement to stabilize

(approximately 4 hours)

• Turn the high humidity trimmer until the readout on the display

is 64.3%.

• Wait for 1 hour and readjust if necessary.

12

3

6

9

4 h

ON/OFF

RANGE

CALIBRATION KIT

Li Cl

11.1%

Na Cl

75.4%

Summary of Contents for HI 9261

Page 1: ...e in Compliance with the CE Directives HI 9161C ON OFF PAPER RANGE RH 5 0 95 0 C 20 0 60 0 INTV instrum nts e UP PRINT DOWN RANGE LOG MICROPROCESSOR PRINTING THERMOHYGROMETER HI 9161 HI 9261 HI 91610...

Page 2: ...by the probe into a digital readout in Relative Humidity The standard cable is 2 meters 7 long and a 5 meter 16 5 version is also available Note the RH probe sensor must never come into contact with w...

Page 3: ...7 5 4 HI 9161C 1 For special applications HI 70606 2 or HI 70606 5 RH probes can also be used These probes come with a sintered cap on the sensor shaft for protection in dusty or corrosive environment...

Page 4: ...0C 1 FUNCTIONAL DESCRIPTION HI 91610 FUNCTIONAL DESCRIPTION HI 91610 FUNCTIONAL DESCRIPTION HI 91610 FUNCTIONAL DESCRIPTION HI 91610 FUNCTIONAL DESCRIPTION HI 91610 FUNCTIONAL DESCRIPTION HI 9261 FUNC...

Page 5: ...power impact type belt 14 characters per line using 38 mm plain paper HI 710034 Printing Intervals Selectable from 1 2 5 10 15 30 60 120 180 min Operating Conditions 0 to 50 C 32 to 122 F max RH 98 no...

Page 6: ...ither the RH or temperature read ing on the display press RANGE The reading will appear without decimal digit when shown on the lower secondary display The tempera ture range on the secondary display...

Page 7: ...r sensors which provide greater accuracy and faster response than conventional thermistor probes Hanna temperature probes are supplied pre calibrated from the factory and are ready to use Available in...

Page 8: ...P or the DOWN keys 24 hour clock To memorize the hour press RANGE once again The minutes will start blinking Likewise use UP and DOWN to set the minutes Press the INTV and the RANGE keys simultaneousl...

Page 9: ...INTV TIME UP DATE DOWN TIME INTV TIME DATE UP To select the hour press the UP or DOWN keys 24 hour clock To set the hour press the TIME key again and the minutes will blink Use the UP or DOWN keys to...

Page 10: ...5 1 1 5 1 6 5 2 R H 2 5 0 C PRINT To view RH on the primary display press RANGE To view temperature on the primary display when in RH mode press RANGE again If is displayed it indicates that no probe...

Page 11: ...OFF key The instrument will then make one final exit printout The running log number can be reset by simply removing the batteries 1 0 0 1 9 6 0 1 0 0 0 5 M 0 0 0 6 1 2 0 1 6 5 1 R H 2 5 0 C ON OFF PR...

Page 12: ...he LOG key The log number and remaining page numbers will appear on display LOG Press the ALT and LOG key simultaneously to enter the logging mode The current log number and the remaining page numbers...

Page 13: ...ns out the meter will not advise the operator and the printouts could be lost If this hap pens stop the printer by pressing ALT and PAPER key simultaneously Data will be kept in memory Insert a new pa...

Page 14: ...ON KIT Li Cl 11 1 Na Cl 75 4 CALIBRATION KIT H 2 O 12cc N a C l Li Cl 11 1 Na Cl 75 4 CALIBRATION KIT 12 3 6 9 4 h CALIBRATION KIT Li Cl 11 1 Na Cl 75 4 12 3 6 9 4 h All Hanna thermohygrometers have b...

Page 15: ...storage or elabora tion without having to connect and disconnect cables between the meter and your PC Data can be further elaborated with the new optional HI 92000 Windows compatible application soft...

Page 16: ...med reset the printer see page 32 SELF DIAGNOSTIC FUNCTIONS SELF DIAGNOSTIC FUNCTIONS SELF DIAGNOSTIC FUNCTIONS SELF DIAGNOSTIC FUNCTIONS SELF DIAGNOSTIC FUNCTIONS Capacity 8000 data samples which are...

Page 17: ...tion HI 9161 HI 9261 and HI 91610 can also be used with other adapters supplying a 12VDC output In this case remember to check the correct polarity of your adapter before connecting it to the meter BA...

Page 18: ...EMPERATURE PROBES These probes use highly sensitive thermistor sensors which provide greater accuracy faster response and a temperature range that is superior to conventional probes All Hanna temperat...

Page 19: ...te of purchase All rights are reserved Reproduction in whole or in part is prohibited without the written consent of the copyright owner Hanna Instruments Inc 584 Park East Drive Woonsocket Rhode Isla...

Page 20: ...ons IEC 801 2 Electrostatic Discharge IEC 801 3 RF Radiated IEC 801 4 Fast Transient EN 55022 Radiated Class B Date of Issue 01 02 1996 D Volpato Engineering Manager On behalf of Hanna Instruments S r...

Page 21: ...A LITERATURE HANNA LITERATURE LAB RECORDING WATER ANALYSIS ENVIROCARE GENERAL CATALOG These and many others catalogs handbooks and leaflets are avail able from Hanna To receive your free copy contact...

Reviews: