6
SPECIFICATIONS
HI 9017
Range
pH
0.00 to 14.00
ISE
± 399.9 mV
ORP
±1999 mV
°C
-10.0 to 125.0
Resolution
pH
0.01
ISE
0.1 mV
ORP
1 mV
°C
0.1
Accuracy
pH
±0.01
(
@
20°C/68°F) ISE
±0.2 mV
ORP
±1 mV
°C
±0.5
Typical EMC pH
±0.02
Deviation
ISE
±0.2 mV
ORP
±1 mV
°C
±0.5
Calibration
2 point with automatic recognition of 3
buffers (pH 4.01, 7.01, 10.01)
Temperature
-10° to +125°C (14 to 257 °F) completely
automatic
with
HI7669/2W temperature
probe;
manual
by means of keyboard
Comp. Interface
RS232C
Electrode
HI1131B
Glass body combination
pH electrode (
included
)
Temperature Probe
HI7669/2W
Temperature probe (
included
)
Input Impedance
10
12
Ohm
Power
Power socket for 12VDC
Environment
0 to 50°C (32 to 122°F); RH 95%
Dimensions
230 x 170 x 70mm (9.1 x 6.7 x 2.7")
Weight
1.3 Kg (2.9 lb.); kit w/holder 3.2 Kg (7 lb.)
ON
key
to switch the instrument on.
OFF
key
to switch the instrument off.
RANGE
key to switch range from pH to mV or vice versa.
CAL
key
used in calibration with standard pH buffers (pH
4.01, 7.01, 10.01)
CON
key
confirmation of calibration data
+/-
key
entering negative values
TEMP
key
used for manual temperature adjustment
0 to 9
keys
numeric and Decimal keys for numeric data input