41
5
PGM
PROGRAM # 4 (HI9318)
This program logs and prints pH/mV and temperature values at a
preset interval between 1 minute and 24 hours.
To enter the program press the 4
key and the PGM key.
The printout defines functions
A, B and C:
A
is a start time,
B
is the time interval,
C
is the finish time.
Hours and minutes must be
separated by a decimal point.
The minimum time interval is
1 minute.
E.g. Starting time 10:20, finishing time 11:30 with a printout interval
of 5 minutes.
When the instrument is in this mode and it is desired to stop before
the end time, press the 0 key plus the PGM key.
PROGRAM #5 (HI9318)
This is the same as program 4 except that function "C" allows you to
select an interrupt value for pH/mV instead of end time.
When this interrupt value is exceeded, the program prints the hour and
minute of the endpoint.
To enter the program press the 5
key plus the PGM key.
PROGRAM 4
A enter start
time
B enter time int-
erval
C enter end time
A
C
B
1
1
0
0
1
.
.
.
2
3
0
0
0
5
4
PGM
1)
2)
3)