15
7
+/-
3
.
TEMP
ºC
DOWN
ºC
UP
°C
°C
With negative temperature values (e.g. -7.3°C) use the +/- key
• press 7
• press decimal point
• press 3
• press +/-
• press TEMP
Notes:
1)
For negative value inputs, always key in the value followed by
the +/- key.
2)
If the +/- key is pressed twice it will toggle back to the
positive value.
The value will be displayed on the left LCD and "°C" symbol will be
flashing.
FOR HI 9218 AND HI 9219 ONLY
Use the UP or DOWN keys to manually adjust the display reading to
the value of the
ChecktempC
or the glass thermometer (e.g. 20°C).
The value will be displayed on the left LCD and "°C" symbol will be
flashing.
FOR HI 9318 ONLY
Press the +/- key (for negative figures) followed by the noted
temperature on the numeric keypad (e.g. 20°C). The value will be
displayed on the small LCD. Then press the TEMP. key and the
manually set temperature will be displayed on the left LCD with the
"°C" symbol blinking.
Note:
with the temperature probe connected, the "°C" symbol is
displayed continuously and it is not possible to manually set
the temperature.
2
0
TEMP.
ºC
SAMPLE
DATE
°C