32
At this stage, the left display shows "P2-1", and the right display
shows the measured pH value.
• press 9
• press decimal point
• press 3
• press 2
• press ENTER A
The left display then shows "P2-2".
• press 6
• press decimal point
• press 0
• press 3
• press ENTER B.
PROGRAM # 3 (HI 9017)
Setting of upper and lower mV alarm limits.
When either limit is reached, the alarm
will be activated and the value will flash
simultaneously.
When the CLEAR key is pressed, the alarm
will be deactivated.
For negative values:
e.g. -100.9 mV and -500mV, -100.9 mV should be treated as larger
than -500mV.
To stop:
• press 0
• press PGM.
ENTER
A
9
.
3
2
pH
ENTER
B
6
.
0
3
pH
ºC
°C
PGM
0
mV
2
PGM
pH