7
6
OPERATIONAL GUIDE
INITIAL PREPARATION
Remove the battery cover on
the rear of your thermom-
eter, unwrap the battery and
connect it to the clip. Insert
the battery into the compart-
ment and replace the cover.
To switch on, press ON/OFF
on the front of the unit.
The thermometer performs a
self diagnostic test. The LCD
dispalys all the segments at
once, to ensure none are
missing.
After the test, the thermometer will switch to
the measurement mode.
If a temperature probe is plugged in, the
unit will display the temperature.
If the probe is not plugged
in, the display will show
"
----
".
To switch your thermometer off, press ON/
OFF again.
MEASURING SCALE (°C/°F) (HI 9057
only)
When the meter is turned on, it automati-
cally defaults to the °C scale. Measurements
can be performed on either the Centigrade
or the Fahrenheit scale.
BAT
HOLD K J T
ºC
ºF
HI
LO
ON
OFF
1. Liquid Crystal Display
2. ON/OFF Key
3. Reading Memory/Hold Key
4. Thermocouple Type Selection Key (K or J or T)
5. Minimum Temperature Recalling Key
6. Maximum Temperature Recalling Key
7. Measuring Scale key (°C or °F).
Specifications
Range
K
-200.0 to 1370°C / -300.0 to 2500°F
J
-200.0 to 900.0°C / -300.0 to 1650°F
T
-200.0 to 400.0°C / -300.0 to 750.0°F
Resolution
0.1°C (up to 999.9°C) 0.1°F(up to 999.9°F)
1°C (over 1000°C) 1°F (over 1000°F)
Accuracy
K=
±0.2% F.S.
(@20°C / 68°F)
J =
±0.2% F.S.
T =
±0.5% F.S.
for one year, excluding probe error
Typical EMC
±3°C
Deviation
±6°F
Probe
K, J and T thermocouple (optional)
(for K-thermocouple see pages 14-17)
Battery
9V/300 hours of continuous use
Environment
0 to 50°C (32 to 122°F); 95% RH
Dimensions
180 x 83 x 40 mm (7.1 x 3.3 x 1.6")
Weight
350 g (12.4 oz.)
FUNCTIONAL DESCRIPTION AND
SPECIFICATIONS OF HI 9057