background image

5

4

1. Liquid Crystal Display
2. ON/OFF Key
3. Reading Memory/Hold Key
4. Measurement Resolution Key (0.1 or 1)
5. Input Probe Selection Key (T1 or T2)
6. Temperatures Difference Key
7. Memory Recalling Key.

Specifications

Range

-200.0 to 1370°C

Resolution

0.1°C (-200.0 to 999.9°C)

1°C (1000 to 1370°C)

Accuracy

±0.2% F.S.

(@20°C / 68°F)

for  one  year, excluding probe error

Typical EMC

±3°C

Deviation

±6°F

Probe

K-thermocouple (optional)

(see pages 14-17)

Battery

9V/300 hours of continuous use

Environment

0 to 50°C (32 to 122°F); 95% RH

Dimensions

180 x 83 x 40 mm (7.1 x 3.3 x 1.6")

Weight

350 g (12.4 oz.)

FUNCTIONAL DESCRIPTION AND

SPECIFICATIONS FOR HI 9056

Standard features include:

• Interchangeable probes

• Clear Liquid Crystal Display

• Two year Warranty

• Possibility of switching between the Cen-

tigrade and the Fahrenheit scale (for

HI 9057

 only)

• HOLD function that freezes the reading

on the display (for 

HI 9057 

only)

• HI-LO memory recall function (for 

HI  9057

only)

• Low battery detection.

All these functions are easily accessible
through the keypad.

Self-explanatory symbols are used to re-
mind the user of the operating mode or
condition.

Hanna manufactures several K-type ther-
mocouple probes for different applications.
See pages 14-17 for more details.

Summary of Contents for HI 9057

Page 1: ...Instruction Manual HI 9056 HI 9057 Portable Microprocessor Thermocouple Thermometers Manufacturers since 1978 These Instruments are in Compliance with the CE Directives...

Page 2: ...9056 5 Functional Description Specifications of HI 9057 6 Operational Guide 7 Display Codes Guide 11 Calibration 12 Battery Replacement 12 Shockproof Rubber Boots 13 Accessories 14 Warranty 18 CE Decl...

Page 3: ...sions 180 x 83 x 40 mm 7 1 x 3 3 x 1 6 Weight 350 g 12 4 oz FUNCTIONAL DESCRIPTION AND SPECIFICATIONS FOR HI 9056 Standard features include Interchangeable probes Clear Liquid Crystal Display Two year...

Page 4: ...the C scale Measurements can be performed on either the Centigrade or the Fahrenheit scale BAT HOLD K J T C F HI LO ON OFF 1 Liquid Crystal Display 2 ON OFF Key 3 Reading Memory Hold Key 4 Thermocoup...

Page 5: ...OLD To change the scale press C F once MEASUREMENT RESOLUTION When switched on HI 9056 and HI 9057 default to a 0 1 digit resolution see specifi cations These instruments automatically change the reso...

Page 6: ...reading in the Centigrade scale F The instrument is reading in the Fahrenheit scale HI Displays the highest measured temperature LO Displays the lowest measured temperature HOLD Freezes current readin...

Page 7: ...alibrated at least once a year For an accurate annual recalibration con tact your nearest Hanna Service Center BATTERY REPLACEMENT When the battery is low the user is warned by the BAT indication on t...

Page 8: ...66B Surface probe max 650 C 1200 F 420mm 13 3 280mm 11 100mm 4 16 5 400mm 15 7 260mm 10 2 155mm 6 1 3mm 0 12 200mm 7 8 5mm 0 20 100mm 4 280mm 7 8 3mm 0 12 155mm 6 1 250mm 9 8 3mm 0 12 11 200mm 7 9 ACC...

Page 9: ...cessories HI 710002 Soft carrying case HI710009 Blue rubber boot HI710010 Orange rubber boot 300mm 11 8 30mm 1 2 130mm 5 1 8mm 0 3 HI 766B1 90 surface probe max 450 C 840 F HI 766B2 Surface probe max...

Page 10: ...e probes are guaranteed for a period of six months This warranty is limited to repair or replacement free of charge Damage due to accident misuse tampering or lack of prescribed maintenance are not co...

Page 11: ...ations The reference literature currently covers areas such as Water Treatment Process Swimming Pools Agriculture Food Laboratory Thermometry and many others New reference material is constantly being...

Reviews: