Hanna Instruments HI 84100 Instruction Manual Download Page 8

8

LCD DESCRIPTION

1)   Stability indicator: when the pump calibration is in progress
2) Stirrer active tags
3) Calibration messages
4) Pump active tags
5) Four digit secondary display
6) “Time” tag: when the time is displayed on the secondary display
7) “ppm” tag: when the titration result is displayed on the primary display
8) Four digit and half main display

KEYPAD DESCRIPTION

1) 

  PURGE

 - to start/stop purging (max purging time is 5 min)

2)

   CAL PUMP

 - to enter pump calibration mode

3) 

  START STOP

 - to start/stop titration or pump calibration

4)

   STIR

 - to start/stop the stirrer while in measurement or purging mode

6) ORP Electrode
7) Beaker
8) BNC electrode connector
9) Fuse
10) Power switch
11) Power cable connector
12) Peristaltic pump

13

• Immerse the ORP electrode approximately 2 cm

(0.8”) into the sample to be tested while paying
attention to not touch the stir bar.

• Insert the dosing tip in the appropriate holder place

and pay attention to not be immersed into solution.

• Press the START STOP button to start the titration.

The display will show “titr” during titration, along
with stirrer and pump tags blinking on the LCD.

• At the end of the titration, the Sulphur Dioxide

concentration is displayed in mg/L (ppm).

Note:

If the equivalence point is not reached or it is not
recognized because of the noisy solution, an error
message will be displayed.

• Fill the 50 mL beaker up to the 50 mL mark with

the wine sample, place the stir bar into the beaker
and put the beaker in the appropriate place on the
minititrator top.

• Fill the 20 mL beaker up to the 5 mL mark with the

HI 84100-51

 Alkaline Reagent and add the content

to the 50 mL beaker.

• Swirl the beaker and wait for 15 minutes.

• Fill the 20 mL beaker up to the 5 mL mark with the

HI 84100-52

 Acid Reagent and add the content to

the 50 mL beaker.

TOTAL SO

2

 MEASUREMENT PROCEDURE

Summary of Contents for HI 84100

Page 1: ...851 9129 0 Fax 07851 9129 99 Greece Tel 210 823 5192 Fax 210 884 0210 Indonesia Tel 21 4584 2941 Fax 21 4584 2942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malays...

Page 2: ...ACCESSORIES 18 WARRANTY 19 19 Hanna Instruments reserves the right to modify the design construction and appearance of its products without advance notice HI 84100 is warranted for two years against...

Page 3: ...ffective interface for the user By simply pressing the START STOP button the instrument will automatically make the titration up to the equivalence point The result is immediately displayed in conveni...

Page 4: ...ted by the human senses is about 2 0 ppm This is also the level which is needed for maximum protection of your wine This is particularly true in the case of sweet and most notably botrytised wines The...

Page 5: ...PREPARATION PROCEDURE Remove the protective cap of the ORP electrode HI 3148B DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT This is normal with electrodes They will disappear when rinsed with water D...

Page 6: ...volume has to be precisely known 15 PUMP CALIBRATION PROCEDURE Warning The calibration of the pump must be performed each time the pump tube the reagent bottle or the pH electrode is changed It is rec...

Page 7: ...ng clockwise Immerse the ORP electrode approximately 2 cm 0 8 into the sample to be tested while paying attention to not touch the stir bar Insert the dosing tip in the appropriate holder place and pa...

Page 8: ...e paying attention to not touch the stir bar Insert the dosing tip in the appropriate holder place and pay attention to not be immersed into solution Press the START STOP button to start the titration...

Page 9: ...ed during testing to ensure best accuracy Purge the peristaltic pump to have fresh titrant when starting a new analysis or calibration Calibrate the peristaltic pump before performing an analysis Anal...

Page 10: ...urging mode message PUMP CALIBRATION MESSAGES 11 This error message appears when the sample concentration exceeds 400 ppm The used standard solution is wrong This error message appears when the input...

Reviews: