An important reason for adding SO

is to avoid oxidation. When there is oxygen around, SO

2

 itself

becomes oxidized before phenol compounds in the wine , and so acts as an oxygen scavenger. Also
SO

suppresses the activity of enzymes that cause browning and other problems.

What is really protecting your wine is molecular SO

2

. When you add SO

2

, depending of circum-

stances, some of it immediately becomes bound. The relationship between the amount of added SO

2

and the amount of SO

2

 remaining free is complex. It is clear, however, that it is largely governed by

the total SO

2

 content of the wine. The rate of binding decreases as the free SO

2

 concentration

increases. The exact relationship between free and bound (total - free) SO

2

 will vary from wine to

wine.
Below 30-60 ppm, 33% to 50% of SO

2

 addition becomes bounded. What remains is called “free”

and it is divided in two parts. The larger, and relatively ineffective free part is “bisulphite” (HSO

3-

). The

smaller part of the free is the active molecular SO

2

. The amount of molecular SO

2

 in your wine

depends both on the level of free SO

2

 present as well as pH. For instance, at pH 3.2, the amount of

free SO

2

 for 0.8 ppm molecular SO

2

 is 22 ppm. At pH 3.5, you will need 43 ppm free - essentially

double.
Free SO

2

 concentration (ppm) for 0.8 ppm molecular SO

2

:

In most situations, 0.8 ppm molecular SO

2

 during bulk storage and at bottling will provide you with

adequate protection from oxidation and bacterial action. This includes prevention of malolactic
bacteria as well.
It is important to remember that the amount of free SO

2

 in the wine depends on three things: how

much is added, how much was present before the addition and how much of your addition promptly
becomes bound.
The level at which molecular SO

2

 can be detected by the human senses is about 2.0 ppm. This is also

the level which is needed for maximum protection of your wine. This is particularly true in the case
of sweet, and most notably, botrytised wines.
The Hanna HI 84100 offers the possibility to test free or total SO

2

 in all the wines including the red

ones, that are difficult to test with manual methods because the color changes are hardly seen.

SIGNIFICANCE OF USE

pH

3.0

3.1

3.2

3.3

3.4

3.5

3.6

3.7

3.8

3.9

Free SO

2

14

18

22

28

35

44

55

69

87

109

17

STORAGE PROCEDURE
To minimize clogging and assure a quick response time, the glass bulb and the junction of the
electrode should be kept moist and not allowed to dry out.
Replace the solution in the protective cap with a few drops of 

HI 70300

 Storage Solution or, in its

absence, Filling Solution (

HI 7082

). Follow the Preparation Procedure before taking measurements.

Note:

NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED WATER.

PERIODIC MAINTENANCE
Inspect the electrode and the cable. The cable used for connection to the instrument must be
intact and there must be no points of broken insulation on the cable or cracks on the electrode
stem or bulb. Connectors must be perfectly clean and dry. If any scratches or cracks are present,
replace the electrode. Rinse off any salt deposits with water.

Probe Maintenance

Refill the reference chamber with fresh electrolyte (

HI 7082

). Allow the electrode to stand

upright for 1 hour. Follow the Storage Procedure above.
CLEANING PROCEDURE
• 

Wine deposits

Soak in Hanna 

HI 70635

 cleaning solution for 15 minutes

• 

Wine stains

Soak in Hanna 

HI 70636

 cleaning solution for 15 minutes

IMPORTANT:

 After performing any of the cleaning procedures, rinse the electrode thoroughly

with distilled water, refill the reference chamber with fresh electrolyte and soak the electrode in

HI 70300

 Storage Solution for at least 1 hour before taking measurements.

For faster response, unscrew the fill hole screw during measurements.

Summary of Contents for HI 84100

Page 1: ...851 9129 0 Fax 07851 9129 99 Greece Tel 210 823 5192 Fax 210 884 0210 Indonesia Tel 21 4584 2941 Fax 21 4584 2942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malays...

Page 2: ...ACCESSORIES 18 WARRANTY 19 19 Hanna Instruments reserves the right to modify the design construction and appearance of its products without advance notice HI 84100 is warranted for two years against...

Page 3: ...ffective interface for the user By simply pressing the START STOP button the instrument will automatically make the titration up to the equivalence point The result is immediately displayed in conveni...

Page 4: ...ted by the human senses is about 2 0 ppm This is also the level which is needed for maximum protection of your wine This is particularly true in the case of sweet and most notably botrytised wines The...

Page 5: ...PREPARATION PROCEDURE Remove the protective cap of the ORP electrode HI 3148B DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT This is normal with electrodes They will disappear when rinsed with water D...

Page 6: ...volume has to be precisely known 15 PUMP CALIBRATION PROCEDURE Warning The calibration of the pump must be performed each time the pump tube the reagent bottle or the pH electrode is changed It is rec...

Page 7: ...ng clockwise Immerse the ORP electrode approximately 2 cm 0 8 into the sample to be tested while paying attention to not touch the stir bar Insert the dosing tip in the appropriate holder place and pa...

Page 8: ...e paying attention to not touch the stir bar Insert the dosing tip in the appropriate holder place and pay attention to not be immersed into solution Press the START STOP button to start the titration...

Page 9: ...ed during testing to ensure best accuracy Purge the peristaltic pump to have fresh titrant when starting a new analysis or calibration Calibrate the peristaltic pump before performing an analysis Anal...

Page 10: ...urging mode message PUMP CALIBRATION MESSAGES 11 This error message appears when the sample concentration exceeds 400 ppm The used standard solution is wrong This error message appears when the input...

Reviews: