background image

10

11

Pay attention to push the cuvette completely down in the holder.

• In order to avoid reagent leaking and to obtain more accurate measure-

ments, close the cuvette first with the supplied HDPE plastic stopper 
and then the black cap.

• Whenever the cuvette is placed into the measurement cell, it must be dry

outside, and free of fingerprints, oil or dirt. Wipe it thoroughly with

HI 731318

  or a lint-free cloth prior to insertion.

• Shaking the cuvette can generate bubbles in the sample, causing higher

readings. To obtain accurate measurements, remove such bubbles by swirling

or by gently tapping the cuvette.

• Do not let the reacted sample stand too long after reagent is added, or

accuracy will be lost.

• It is possible to take multiple readings in a row, but it is recommended to

take a new zero reading for each sample and to use the same cuvette for

zeroing and measurement when possible (for most precise results follow the
measurement procedures carefully).

• Discard the sample immediately after the reading is taken, or the glass

might become permanently stained.

• All the reaction times reported in this manual are at 20°C (68°F). As a general rule of thumb, they

should be doubled at 10°C (50°F) and halved at 30°C (86°F).

REAGENT BLANK CORRECTION

• Some methods require a “reagent blank correction”. The blank and the sample are prepared exactly

in the same way, only the blank is deionized water instead of sample.

A blank cuvette may be used more than once: stability and storing conditions are described for each

method in the related chapter.

INTERFERENCES

• In the method measurement section the most common interferences that may be present in an average

wastewater matrix have been reported. It may be that for a particular treatment process other

compounds do interfere with the method of analysis.

METHOD REFERENCE TABLE

HI 83206 - ENVIRONMENTAL TESTING

Method

Method

Page

description

1

Ammonia MR

17

2

Ammonia LR

19

3

Free Chlorine

21

4

Total Chlorine

24

5

Chromium VI HR

27

6

Chromium VI LR

29

7

Color of Water

31

8

Copper HR

33

9

Copper LR

35

10

Cyanuric Acid

37

11

Molybdenum

39

12

Nickel HR

42

Method

Method

Page

description

13

Nickel LR

44

14

Nitrate

47

15

Nitrite HR

49

16

Nitrite LR

51

17

Dissolved Oxygen

53

18

pH

55

19

Phosphate HR

57

20

Phosphate LR

59

21

Phosphorus

61

22

Silica

63

23

Silver

66

24

Zinc

69

HEALTH & SAFETY

The chemicals contained in the reagent kits may be hazardous if improperly handled. Read the
Material Safety Data Sheet (MSDS) before performing tests.

Safety equipment:  Wear suitable eye protection and clothing when required, and follow instructions carefully.

Reagent spills: If a reagent spill occurs, wipe up immediately and rinse with plenty of water.
If reagent contacts skin, rinse the affected area thoroughly with water. Avoid breathing released vapors.

Summary of Contents for HI 83206

Page 1: ...1 Instruction Manual HI 83206 Multiparameter Bench Photometer for Environmental Testing w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives...

Page 2: ...tible Software helps users manage all their results Dear Customer Thank you for choosing a Hanna product Please read this instruction manual carefully before using the instrument This manual will prov...

Page 3: ...are optically identical to provide the same measurement conditions Most of methods use the same cuvette for both so it is important that measurements take place in the same optical point The instrumen...

Page 4: ...the select method menu Press to move up in a menu or a help screen to increment a set value to access second level functions Press to move down in a menu or a help screen to decrement a set value to a...

Page 5: ...mounts may cause interference they should be removed by treatment with active carbon and filtration Ensure the cuvette is filled correctly the liquid in the cuvette forms a convexity on the top the bo...

Page 6: ...The blank and the sample are prepared exactly in the same way only the blank is deionized water instead of sample A blank cuvette may be used more than once stability and storing conditions are descri...

Page 7: ...g and deleting You can view and delete the data log by pressing the RCL key Deleting is based on the LIFO last in first out scheme Additionally you can delete the data records all at once CHEMICAL FOR...

Page 8: ...enabled a short beep is heard every time a key is pressed A long beep alert sounds when the pressed key is not active or an error condition is detected Press the functional key to enable disable the...

Page 9: ...eractive contextual help mode that assists the user at any time To access help screens pressHELP The instrument will display additional information related to the current screen To read all available...

Page 10: ...ide various aliphatic and aromatic amines Ammonia LR SPECIFICATIONS Range 0 00 to 3 00 mg L Resolution 0 01 mg L Accuracy 0 04 mg L 4 of reading Typical EMC 0 01 mg L Deviation Light Source Tungsten l...

Page 11: ...ette with 10 mL of unreacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Free Chlorine FREE CHLORINE 10 mL Ammonia LR Remove the cuvette Add 4 drops o...

Page 12: ...ey The meter will show 0 0 when the meter is zeroed and ready for measurement Remove the cuvette Powder reagents procedure Add the content of one packet of HI 93701 0 DPD reagent Replace the cap and s...

Page 13: ...instrument x 3 x 3 x 1 Total Chlorine TOTAL CHLORINE SPECIFICATIONS Range 0 00 to 3 50 mg L Resolution 0 01 mg L Accuracy 0 03 mg L 3 of reading Typical EMC 0 01 mg L Deviation Light Source Tungsten l...

Page 14: ...will show 0 0 when the meter is zeroed and ready for measurement Remove the cuvette and add the content of one packet of HI 93723 0 reagent Replace the cap and shake vigorously for about 10 seconds 1...

Page 15: ...the cap Place the cuvette into the holder and close the lid Press ZERO key The meter will show 0 0 when the meter is zeroed and ready for measurement Remove the cuvette and add the content of one pac...

Page 16: ...ce the blank 1 into the holder and close the lid Press ZERO key The meter will show 0 0 when the meter is zeroed and ready for measurement Remove the blank 10 mL 1 1 Reinsert the cuvette into the inst...

Page 17: ...te with 10 mL of unreacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The meter will show 0 0 when the meter is zeroed and ready for m...

Page 18: ...ion section see page 12 Fill the cuvette with 10 mL of unreacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The meter will show 0 0 wh...

Page 19: ...ed sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The meter will show 0 0 when the meter is zeroed and ready for measurement Fill a gradua...

Page 20: ...other accessories see page 73 MEASUREMENT PROCEDURE Select the Molybdenum method using the procedure described in the Method Selection section see page 12 Fill the cuvette with 10 mL of unreacted sam...

Page 21: ...above 10 mg L Iron above 50 mg L Nickel above 50 mg L Nitrite as NO2 Sulfate above 200 mg L Highly buffered samples or with extreme pH may exceed the buffering capacity of the reagents Fill one gradua...

Page 22: ...he reaction between nickel and the reagent causes a blue tint in the sample REQUIRED REAGENTS Code Description Quantity HI 93726 0 Powder reagent 1 packet REAGENT SETS HI 93726 01 Reagents for 100 tes...

Page 23: ...ical EMC 0 001 mg L Deviation Light Source Tungsten lamp with narrow band interference filter 575 nm Method Adaptation of the PAN method The reaction between nickel and the reagents causes an orange t...

Page 24: ...cedure described in the Method Selection section see page 12 Using the pipette fill the cuvette with 6 ml of sample up to half of its height and replace the cap Place the cuvette into the holder and c...

Page 25: ...f one packet of HI 93708 0 reagent Replace the cap and shake gently until completely dissolved 10 mL Nitrate Replace the cap and immediately shake vigorously up and down for exactly 10 seconds Continu...

Page 26: ...ark with 10 mL of unreacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The display will show 0 0 when the meter is zeroed and ready fo...

Page 27: ...93732B 0 Add more sample to fill the bottle completely Replace the cap again and ensure that a part of the sample spills over This is to make sure that no air bubbles have been trapped inside which co...

Page 28: ...cuvette and add 5 drops of HI 93710 0 Phenol Red Indicator Replace the cap and mix the solution 10 mL x 5 Dissolved Oxygen Let the sample stand and the flocculent agent will start to settle After app...

Page 29: ...3717B 0 Reagent B 1 packet REAGENT SETS HI 93717 01 Reagents for 100 tests HI 93717 03 Reagents for 300 tests For other accessories see page 73 MEASUREMENT PROCEDURE Select the Phosphate HR method usi...

Page 30: ...Press ZERO key The display will show 0 0 when the meter is zeroed and ready for measurement Remove the cuvette and add the content of one packet of HI 93713 0 reagent Replace the cap and shake gently...

Page 31: ...reacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The display will show 0 0 when the meter is zeroed and ready for measurement Remove...

Page 32: ...eagent Replace the cap and swirl the solution Wait for 4 minutes add the content of one packet of HI 93705B 0 Citric acid reagent and shake until it is completely dissolved Wait for 1 minute This is t...

Page 33: ...e the cuvette and add the content of one packet of HI 93705C 0 Amino acid reagent and shake until it is completely dissolved Reinsert the cuvette into the instrument Press TIMER and the display will s...

Page 34: ...uracy 0 005 mg L 10 of reading Typical EMC 0 001 mg L Deviation Light Source Tungsten lamp with narrow band interference filter 575 nm Method Adaptation of the PAN method The reaction between silver a...

Page 35: ...reagent close the cylinder and invert several times to mix until completely dissolved Fill one cuvette with 10 mL of the reacted sample up to the mark 20 mL 10 mL Silver Press ZERO key The display wil...

Page 36: ...be caused by Aluminum above 6 mg L Cadmium above 0 5 mg L Copper above 5 mg L Iron above 7 mg L Manganese above 5 mg L Nickel above 5 mg L Zinc Place the cap and insert the cuvette into the instrumen...

Page 37: ...Zincon ACCESSORIES REAGENT SETS HI 93700 01 100 ammonia LR tests HI 93700 03 300 ammonia LR tests HI 93701 01 100 free chlorine tests powder HI 93701 03 300 free chlorine tests powder HI 93701 F 300...

Page 38: ...r dealer or the Hanna Customer Service Center nearest to you To find the Hanna Office in your vicinity check our home page at www hannainst com RecommendationsforUsers Before using these products make...

Page 39: ...567 e mail sales hanna co jp Korea Tel 02 2278 5147 Fax 02 2264 1729 e mail mccoyhan chollian net Malaysia Tel 603 5638 9940 Fax 603 5638 9829 e mail hannamal tm net my Norway Tel 23 3811 00 Fax 23 38...

Reviews: