Molybdenum

MOLYBDENUM

SPECIFICATIONS

Range

0.0 to 40.0 mg/L

Resolution

0.1 mg/L

Accuracy

±0.3 mg/L ±5% of reading

Typical EMC

±0.1 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 420 nm

Method

Adaptation of the mercaptoacetic acid method. The reaction between molybdenum and
the reagents causes a yellow tint in the sample.

REQUIRED REAGENT

Code

Description

Quantity

HI  93730

A

-0

Reagent A

1 packet

HI  93730

B

-0

Reagent B

1 packet

HI  93730

C

-0

Reagent C

1 packet

REAGENT SETS

HI 93730-01

Reagents for 100 tests

HI 93730-03

Reagents for 300 tests

For other accessories see page 73.

MEASUREMENT PROCEDURE

• Select the 

Molybdenum method using the procedure

described in the 

Method Selection section (see page 12).

• Fill the cuvette with 10 mL of unreacted sample (up to

the mark) and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press ZERO key. The display will show “-0.0-” when

the meter is zeroed and ready for measurement.

Cyanuric Acid

• Fill a second cuvette with 10 mL of the reacted sample up to the

mark. Replace the cap.

• Reinsert the cuvette into the instrument.

• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait

for 45 seconds and press READ. When the timer ends the meter will perform the reading.

• The instrument displays concentration in mg/L of cyanuric acid.

10 mL

Summary of Contents for HI 83206

Page 1: ...1 Instruction Manual HI 83206 Multiparameter Bench Photometer for Environmental Testing w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives...

Page 2: ...tible Software helps users manage all their results Dear Customer Thank you for choosing a Hanna product Please read this instruction manual carefully before using the instrument This manual will prov...

Page 3: ...are optically identical to provide the same measurement conditions Most of methods use the same cuvette for both so it is important that measurements take place in the same optical point The instrumen...

Page 4: ...the select method menu Press to move up in a menu or a help screen to increment a set value to access second level functions Press to move down in a menu or a help screen to decrement a set value to a...

Page 5: ...mounts may cause interference they should be removed by treatment with active carbon and filtration Ensure the cuvette is filled correctly the liquid in the cuvette forms a convexity on the top the bo...

Page 6: ...The blank and the sample are prepared exactly in the same way only the blank is deionized water instead of sample A blank cuvette may be used more than once stability and storing conditions are descri...

Page 7: ...g and deleting You can view and delete the data log by pressing the RCL key Deleting is based on the LIFO last in first out scheme Additionally you can delete the data records all at once CHEMICAL FOR...

Page 8: ...enabled a short beep is heard every time a key is pressed A long beep alert sounds when the pressed key is not active or an error condition is detected Press the functional key to enable disable the...

Page 9: ...eractive contextual help mode that assists the user at any time To access help screens pressHELP The instrument will display additional information related to the current screen To read all available...

Page 10: ...ide various aliphatic and aromatic amines Ammonia LR SPECIFICATIONS Range 0 00 to 3 00 mg L Resolution 0 01 mg L Accuracy 0 04 mg L 4 of reading Typical EMC 0 01 mg L Deviation Light Source Tungsten l...

Page 11: ...ette with 10 mL of unreacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Free Chlorine FREE CHLORINE 10 mL Ammonia LR Remove the cuvette Add 4 drops o...

Page 12: ...ey The meter will show 0 0 when the meter is zeroed and ready for measurement Remove the cuvette Powder reagents procedure Add the content of one packet of HI 93701 0 DPD reagent Replace the cap and s...

Page 13: ...instrument x 3 x 3 x 1 Total Chlorine TOTAL CHLORINE SPECIFICATIONS Range 0 00 to 3 50 mg L Resolution 0 01 mg L Accuracy 0 03 mg L 3 of reading Typical EMC 0 01 mg L Deviation Light Source Tungsten l...

Page 14: ...will show 0 0 when the meter is zeroed and ready for measurement Remove the cuvette and add the content of one packet of HI 93723 0 reagent Replace the cap and shake vigorously for about 10 seconds 1...

Page 15: ...the cap Place the cuvette into the holder and close the lid Press ZERO key The meter will show 0 0 when the meter is zeroed and ready for measurement Remove the cuvette and add the content of one pac...

Page 16: ...ce the blank 1 into the holder and close the lid Press ZERO key The meter will show 0 0 when the meter is zeroed and ready for measurement Remove the blank 10 mL 1 1 Reinsert the cuvette into the inst...

Page 17: ...te with 10 mL of unreacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The meter will show 0 0 when the meter is zeroed and ready for m...

Page 18: ...ion section see page 12 Fill the cuvette with 10 mL of unreacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The meter will show 0 0 wh...

Page 19: ...ed sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The meter will show 0 0 when the meter is zeroed and ready for measurement Fill a gradua...

Page 20: ...other accessories see page 73 MEASUREMENT PROCEDURE Select the Molybdenum method using the procedure described in the Method Selection section see page 12 Fill the cuvette with 10 mL of unreacted sam...

Page 21: ...above 10 mg L Iron above 50 mg L Nickel above 50 mg L Nitrite as NO2 Sulfate above 200 mg L Highly buffered samples or with extreme pH may exceed the buffering capacity of the reagents Fill one gradua...

Page 22: ...he reaction between nickel and the reagent causes a blue tint in the sample REQUIRED REAGENTS Code Description Quantity HI 93726 0 Powder reagent 1 packet REAGENT SETS HI 93726 01 Reagents for 100 tes...

Page 23: ...ical EMC 0 001 mg L Deviation Light Source Tungsten lamp with narrow band interference filter 575 nm Method Adaptation of the PAN method The reaction between nickel and the reagents causes an orange t...

Page 24: ...cedure described in the Method Selection section see page 12 Using the pipette fill the cuvette with 6 ml of sample up to half of its height and replace the cap Place the cuvette into the holder and c...

Page 25: ...f one packet of HI 93708 0 reagent Replace the cap and shake gently until completely dissolved 10 mL Nitrate Replace the cap and immediately shake vigorously up and down for exactly 10 seconds Continu...

Page 26: ...ark with 10 mL of unreacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The display will show 0 0 when the meter is zeroed and ready fo...

Page 27: ...93732B 0 Add more sample to fill the bottle completely Replace the cap again and ensure that a part of the sample spills over This is to make sure that no air bubbles have been trapped inside which co...

Page 28: ...cuvette and add 5 drops of HI 93710 0 Phenol Red Indicator Replace the cap and mix the solution 10 mL x 5 Dissolved Oxygen Let the sample stand and the flocculent agent will start to settle After app...

Page 29: ...3717B 0 Reagent B 1 packet REAGENT SETS HI 93717 01 Reagents for 100 tests HI 93717 03 Reagents for 300 tests For other accessories see page 73 MEASUREMENT PROCEDURE Select the Phosphate HR method usi...

Page 30: ...Press ZERO key The display will show 0 0 when the meter is zeroed and ready for measurement Remove the cuvette and add the content of one packet of HI 93713 0 reagent Replace the cap and shake gently...

Page 31: ...reacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Press ZERO key The display will show 0 0 when the meter is zeroed and ready for measurement Remove...

Page 32: ...eagent Replace the cap and swirl the solution Wait for 4 minutes add the content of one packet of HI 93705B 0 Citric acid reagent and shake until it is completely dissolved Wait for 1 minute This is t...

Page 33: ...e the cuvette and add the content of one packet of HI 93705C 0 Amino acid reagent and shake until it is completely dissolved Reinsert the cuvette into the instrument Press TIMER and the display will s...

Page 34: ...uracy 0 005 mg L 10 of reading Typical EMC 0 001 mg L Deviation Light Source Tungsten lamp with narrow band interference filter 575 nm Method Adaptation of the PAN method The reaction between silver a...

Page 35: ...reagent close the cylinder and invert several times to mix until completely dissolved Fill one cuvette with 10 mL of the reacted sample up to the mark 20 mL 10 mL Silver Press ZERO key The display wil...

Page 36: ...be caused by Aluminum above 6 mg L Cadmium above 0 5 mg L Copper above 5 mg L Iron above 7 mg L Manganese above 5 mg L Nickel above 5 mg L Zinc Place the cap and insert the cuvette into the instrumen...

Page 37: ...Zincon ACCESSORIES REAGENT SETS HI 93700 01 100 ammonia LR tests HI 93700 03 300 ammonia LR tests HI 93701 01 100 free chlorine tests powder HI 93701 03 300 free chlorine tests powder HI 93701 F 300...

Page 38: ...r dealer or the Hanna Customer Service Center nearest to you To find the Hanna Office in your vicinity check our home page at www hannainst com RecommendationsforUsers Before using these products make...

Page 39: ...567 e mail sales hanna co jp Korea Tel 02 2278 5147 Fax 02 2264 1729 e mail mccoyhan chollian net Malaysia Tel 603 5638 9940 Fax 603 5638 9829 e mail hannamal tm net my Norway Tel 23 3811 00 Fax 23 38...

Reviews: