INSTRUMENT DESCRIPTION
FUNCTIONAL DESCRIPTION
The measurement process is carried out in two phases: first the meter is zeroed and then the actual
measurement is performed.
The cuvette has a very important role because it is an optical element and thus requires particular attention. It
is important that both the measurement and the calibration (zeroing) cuvette are optically identical to provide the
same measurement conditions. Most methods use the same cuvette for both, so it is important that measurements
are taken at the same optical point. The instrument and the cuvette cap have special marks that must be
aligned in order to obtain better reproducibility.
The surface of the cuvette must be clean and not scratched. This is to avoid measurement interference due to
unwanted reflection and absorption of light. It is recommended not to touch the cuvette walls with hands.
Furthermore, in order to maintain the same conditions during the zeroing and the measurement phases,
it is necessary to cap the cuvette to prevent any contamination.