Ammonia MR
SPECIFICATIONS
Range
0.00 to 10.00 mg/L
Resolution
0.01 mg/L
Accuracy
±0.05 mg/L ±5% of reading at 25 °C
Typical EMC
±0.01 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 420 nm
Method
Adaptation of the
ASTM Manual of Water and Environmental Technology, D1426-92,
Nessler method. The reaction between ammonia and reagents causes a yellow tint in
the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93715
A
-0
First Reagent
4 drops (6 drops for seawater)
HI 93715
B
-0
Second Reagent 4 drops (10 drops for seawater)
REAGENT SETS
HI 93715-01
Reagents for 100 tests
HI 93715-03
Reagents for 300 tests
For other accessories see page 132.
MEASUREMENT PROCEDURE
• Select the
Ammonia MR method using the procedure described in
the
Method Selection section (see page 12).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the
Zero
key. The display will show “
-0.0-
” when the meter is zeroed and ready for
measurement.
AMMONIA MEDIUM RANGE