
124
Silver
• Press the
Zero
key. The display will show “
-0.0-
” when the meter is zeroed and ready for
measurement.
• Fill a second cuvette up to the mark with 10 mL of the reacted sample.
• Add 3 drops of Dispersing Agent (HI 93703-51),
replace the cap and invert gently to mix for about
10 seconds.
• Insert the second cuvette into the instrument.
• Press
Read
to start the reading. The instrument displays the results in
mg/L of silver
.
INTERFERENCES
Interference may be caused by:
Al
3+
above 30 mg/L
Fe
2+
above 1.5 mg/L
Ca
2+
above 1000 mg/L as CaCO
3
Fe
3+
above 10 mg/L
Cd
2+
above 20 mg/L
K
+
above 500 mg/L
Cl¯ above 8000 mg/L
Mn
2+
above 25 mg/L
Co
2+
above 1.5 mg/L
Mg
2+
above 1000 mg/L as CaCO
3
Cr
3+
above 20 mg/L
Na
+
above 5000 mg/L
Cr
6+
above 40 mg/L
Ni
2+
above 1.5 mg/L
Cu
2+
above 15 mg/L
Pb
2+
above 20 mg/L
F¯ above 20 mg/L
Zn
2+
above 30 mg/L
10 mL
# 2
# 2