then S1-H1≥S2+H2, S2≥LA, HA≥S1
;
If M1 = 2 and M2 = 1
then S2-H2≥S1+H1, S1≥LA, HA≥S2;
If M1 = 3 and M2 = 2
then S1≥S2+H2, S2≥LA, HA≥S1+D1;
If M1 = 2 and M2 = 3
then S1+H1≤S2, S1≥LA, HA≥S2+D2;
If M1 = 4 and M2 = 1
then S1≤S2–H2, S1–D1≥LA, HA≥S2;
If M1 = 1 and M2 = 4
then S1–H1≥S2, S2–D2≥LA, HA≥S1;
If M1 = 3 and M2 = 4
then S1≥S2, S2–D2≥LA, HA≥S1+D1;
If M1 = 4 and M2 = 3
then S2≥S1, S1–D1≥LA, HA≥S2+D2;
where the minimum deviation (D1 or D2) is 0.5% of the maximum range value.
Note
When a wrong setup value is confirmed, the controller does
not skip to the next setup item but remains in the current item
displaying a flashing “ERROR” indicator until the parameter
value is changed by the user (the same is also true for the
setup code selection).
Note
In some circumstances, the user cannot succeed in setting a parameter to a desired value
if the related parameters are not changed beforehand; e.g. to set a EC high setpoint to
10.0 mS the high alarm must be set to a value greater than 10.0 mS first.