SETUP MODE
HI700
and
HI710
offer a multitude of possibilities from ON/OFF or PID dosage to
analog recorder output and from alarm to selftest features.
The Setup Mode allows the user to set all needed characteristics of the meter.
The setup mode is entered by pressing SETUP and entering the password when the
device is in idle or control mode.
Generally speaking, if the password is not inserted the user can only
view the setup parameters (except for password) without modifying
them (and the device remains in control mode). An exception is
certain setup items, or flags, which can activate special tasks when
set and confirmed.
Each setup parameter (or setup item) is assigned a two-digit setup code which is entered
and displayed on the secondary LCD.
The setup codes can be selected after password and CFM are pressed. When CFM is
pressed, the current setup item is saved on EEPROM and the following item is displayed.
Whenever LCD is pressed, the device reverts back
to control mode. The same is true when CFM is
pressed on the last setup item.
The possible transitions in setup mode are the
following:
ENTERING THE PASSWORD
• Press SETUP to enter the setup mode. The LCD will display “0000” on the upper part
and “PAS” on the lower. The first digit of the upper part of the LCD will blink.
• Enter the first value of the password by
the or keys.
• Then confirm the displayed digit with and move to the next
one.
• When the whole password has been inserted, press CFM to confirm
it
.