Code
Valid Values
Default
PW
00 Factory ID
0 to 9999
0000
no
01 Process ID
0 to 99
00
no
02 Control enable/
disable
0: C.M. disabled
1: C.M. enabled
0
no
03 Range
(depends on mode)
1: 0.0-199.9 µS (or 100.0 ppm)
2: 0-1999 µS (or 1000 ppm)
3: 0.00-19.99 mS (or 10.00 ppt)
4: 0.0-199.9 mS (or 100.0 ppt)
4
no
04 Reference Temperature 20 °C or 25 °C
25°C
no
05 Temperature Coefficient 0.00 to 10.00 %/ °C
2.00
no
06 Input Selection
0: conductivity probe
1: 4-20 mA input signal
0
no
07 Temperature
compensation
ATC: Automatic
User: Manual
ATC
no
08 TDS Factor (HI710
only)
0.00 to 1.00
0.50
no
11 Relay 1 mode
(M1)
0: disabled
1: ON-OFF high setpoint
2: ON-OFF low setpoint
3: PID, high setpoint
4: PID, low setpoint
0
no
12 Relay 1 setpoint (S1)
0.5 to 99.5% full scale
25% f.s.
no
13 Relay 1 hysteresis (H1) 0 to 5% f.s.
1% f.s.
no
14 Relay 1 deviation (D1) 0.5 to 10% f.s.
1% f.s.
no
15 Relay 1 reset time
0.1 to 999.9 minutes
999.9
no
16 Relay 1 rate time
0.0 to 999.9 minutes
0.0
no
21 Relay 2 mode (M2)
same as relay 1
0
no
22 Relay 2 setpoint (S2)
0.5 to 99.5% full scale
75% f.s.
no
23 Relay 2 hysteresis (H2) 0 to 5% f.s.
1% f.s.
no
24 Relay 2 deviation (D2) 0.5 to 10% f.s.
1% f.s.
no
25 Relay 2 reset time
0.1 to 999.9 minutes
999.9
no
26 Relay 2 rate time
0.0 to 999.9 minutes
0.0
no