47
SEED CORRECTION
To apply seed corrections to the sample data follow the
following procedure:
• Press
then
to select BOD mode.
• Press
and
to view samples.
Note: This list contains bottles with initial data, with
5 day BOD determinations and 5 day BOD determina-
tions with seed corrections. The dates reflect the last
date the sample was evaluated.
• Press
then
to select the sample and
seed bottles.
• If more than one Seed Bottle is used, the average
value of the seeds will be used for Seed correction.
• Press
to move samples into the lower box.
These bottles will be corrected with the selected seed
values.
BOD MEASUREMENT