Clear Calibration
Accessing this option, the existent conductivity calibration can be cleared. If the calibration is cleared, another
calibration has to be performed.
To clear calibration:
• Press
while in
Conductivity mode.
• Press
.
• Use
or
to highlight the
Calibration option.
• Use
or
to highlight the
Clear Calibration option.
• Press
to clear calibration. A pop-up menu will be displayed asking for confirmation (if calibration is
available).
• Press
to confirm or press
to escape without saving and return to the
Calibration options.
Cell constant
manual editing:
The conductivity probe can also be calibrated by entering the cell constant value.
To edit the cell constant value:
• Press
while in
Conductivity mode.
• Press
.
• Use
or
to highlight the
Cell Constant
option.
• Press
to reset the cell constant value to default (1.0000/
cm).
• Use
/
to increase / decrease the value.
• Press
to confirm the new value or press
to exit
without modifying.
Probe Type
This option allows the user to obtain some information about the connected conductivity probe: name, default cell
constant, range and rings number. Both
HI 76312
and
HI 76313
probes are recognized by the meter.