27
Calibration Points
The user can choose between
Single Point and Multi Points calibration.
To set the calibration points:
• Press
while in
Conductivity mode.
• Press
.
• Use
or
to highlight the
Calibration option.
• Press
and then use
or
to highlight
the
Calibration Points option.
• Press
to choose
Multiple Points calibration.
• Press
to choose
Single Point calibration.
The following options are available for calibration:
Standard Recognition
The user can choose between
Automatic recognition (from 6 Hanna standards available) or User Standard (when
custom standards are used for calibration).
• Press
while in
Conductivity mode.
• Press
.
• Use
or
to highlight the
Calibration option.
• Press
and then use
or
to highlight
the
Standard Recognition option.
• Press
to choose
Automatic recognition mode.
• Press
to choose
User Standard mode.