21
CONDUCTIVITY SETUP
The
Conductivity Setup menu allows the user to set the parameters related to the conductivity measurement
and calibration.
Accessing Conductivity Setup
• Press
while in
Measure mode and then
to
select the
Conductivity measurement mode.
• Press
and then
to access
Conductivity Setup
menu.
To access a conductivity setup options:
• Use
or
to highlight the desired option.
• Press
to access the selected option or
to
exit setup.
The following is a detailed description of the
Conductivity Setup
option screens.
Profile
This option opens the Profile manager. Enabling Profile allows the user to Save, Load or Delete an
application Profile. The Profile option allows the user to store up to ten separate profile applications. Each
Profile can be named and recalled at a moment’s notice. A profile is a sensor setup complete with
measurement units, logging and display preferences, calibration standards (Standards including custom),
setup of the Display screen for measurement (i.e. graphing, GLP) and any other sensor configuration. Once
saved, the exact same profile can be used at another time. This is a handy feature if the meter is used
occasionally for additional applications because it saves time in the setup of the meter and ensures the same
procedure will be used.
To save the measurement configuration for Conductivity mode:
• Press
, then
and use
or
to highlight
Profile option.
• Press
/
to enable / disable this feature.
The available options are:
Profile Feature
: enable or disable the profile feature.
Save Profile
: save the current profile.
Save Profile As...
: save current profile using a specific name.
Load Profile
: load from available profiles.
Delete Profile
: delete a profile.