Error Messages During Measurement
If the pH or temperature exceeds the limits of the sensor, the message
“ELECTRODE OUT OF SPEC” will scroll on the third LCD line. The
temperature will continue to be displayed. If temperature exceeds the
meter specification of 120 ºC, then “120 ºC” will blink on the display.
If interval logging, the message “OUT OF SPEC.” will alternate with
the LOG specific message in both these cases and the Log file will
indicate a “ºC!” next to the data.
In case the temperature sensor is damaged, “BROKEN TEMPERATURE
SENSOR” will be displayed and the temperature will display “25.0”
and the unit tag blinking on the second LCD line. The Log file will
indicate “ºC!!” next to the data.
mV Reading Of The pH
The mV reading of the measured pH can be displayed on the LCD by
pressing the RANGE/
u
key.
Range Specific Messages Displayed On Third LCD Line During
Measurement
All the messages described in General Setup (page 14) are displayed
in pH range.
Temperature Sensor Problem (if there is one)
Cal Due or Offset and Slope Value
Time
Date
Battery or Charge Status
Logging Messages
Out of Calibration Range
RELATIVE mV CALIBRATION
Calibration Procedure
Pour small quantity of solution into clean beakers. If possible, use plastic beakers to minimize any
EMC interferences.
OPERA
TIONAL GUIDE