The following General Setup options are displayed regardless of the sensor being used. These settings
remain when switching to another probe type or when no probe is attached. Options are tabulated in
the table below with choices and default values. Options are accessed by pressing
SETUP/CLR
key.
Loop through the options by using the
keys. To modify a setting, press
CAL/MODIFY
key. The
option may be modified by using
RANGE
/
u
,
keys. Press
GLP/CFM
key to confirm the
change. To exit SETUP press
SETUP/CLR
key.
GENERAL SETUP
SETUP/ INST
ALLA
TION
Option
Description
Choices
Default
Basic mode
*Only seen when
cable connection
between micro
USB and PC is
made.
Select if PC is being used for
charging battery (and meter
will be used for logging) or if
Data will be exported to the PC.
LOG ON EDGE or EXPORT
TO PC
LOG ON EDGE
Available
Log
Select log type to be used from
3 types of logging:
Manual log on demand
Manual log on stability (3 types
of stability criteria available)
Timed interval lot logging
Manual Log
Stability Log: Fast,
Medium, Accurate;
Interval Log: Seconds:
5,10, 30;
Interval Log Minutes: 1, 2,
5, 15, 30, 60, 120, 180.
Interval
(5 Sec)
Manual log
or Stability
log: Medium
Set Calibration
Expiration
Warning
Meter will indicate “CAL
DUE” when set time in this
parameter has been exceeded.
1, 2, 3, 4, 5, 6, 7 days
or OFF
7 days
Not available
Probe Specific
Parameters that are specific to a measurement type are inserted here in the SETUP list.
Set Date
Press
CAL/MODIF
Y key to Set
current date, displayed in ISO
format.
Press
GLP/CFM
key to save
changes.
YYYY/MM/DD Date
Set date
Available
Set Time
Press
CAL/MODIF
Y key to Set
current time, displayed in ISO
format.
Press
GLP/CFM
key to save
changes.
24 hr:MM:SS Time
Set time
Available