Logged data on
edge
®
pH
can be transferred from the meter to a USB flash drive by using the log
recall function. The minimum requirement for the drive is USB 2.0. Select the pH record you wish to
export and follow the simple steps below.
Connect USB flash drive to the USB port, located on the top of the
meter. Press the
RCL
key. Select Manual, Stability, or interval lots by
using the
keys. Press the
LOG
key (not
GLP/CFM
). The “USB
HOST”tag should come on.
“PLEASE WAIT” message appears followed by “EXPORT”. Press
GLP/CFM
key to export the selected record or lot. If
GLP/CFM
key is
not pressed in 10 seconds, the USB host will become inactive.
The meter will display the percentage of export.
The export percentage should go to 100%. Remove USB flash drive.
If the selected file is already saved on the flash drive,
edge
®
pH
will
ask for confirmation of overwriting the existing file. The message
“OVERWRITE’ and “CFM” tag will blink. Press
GLP/CFM
key for
overwriting the existing file or
CAL/MODIFY
key to exit without
exporting.
After exporting the display will return to the selected file. Press the
RCL
key twice to return to measurements.
Note: Do not remove USB flash drive during an active export transfer.