25
Calibration Timeout
Calibration timeout feature is used to set a scheduled reminder for calibration or service. It can be
configured 1‑30 days or disabled (no message). A reminder message will be displayed with the
measurement indicating it's time to service the probe.
Mode
Press functional key to change EC with TDS mode.
Press
Escape
to return to the general menu.
TDS Factor
Press functional key to choose desired TDS factor: 0.5 (500ppm)
or 0.7 (700ppm). Press
Escape
to return to the general menu.
5.4.5. TEMPERATURE SETUP
Alarm High
and
Alarm Low
in the Temperature menu have two options
Enable
or
Disable
:
When alarms are set to
Enable
, press
Modify
to adjust the Temperature alarm value.
Use
▲▼
keys to set Temperature alarm value.
Note: The high alarm must be set to a value greater than the low alarm value.