DISSOLVED OXYGEN CALIBRATION MESSAGES
If the reading is outside limits, “WRONG STANDARD“ message will be
displayed.
If the temperature is out of range (0.0 ‑ 50.0 ºC) during calibration, then
the message “WRONG STANDARD TEMPERATURE” will be displayed and
temperature value will blink.
DISSOLVED OXYGEN GLP INFORMATION
GLP refers to a quality control function used to ensure uniformity of probe calibrations and
measurements. The dedicated
GLP/CFM
key opens a file of the latest calibration information. Use
the keys to scroll the stored information. This includes the standards used, temperature of the
standard, Altitude and Salinity factors, time and date of the last calibration, the expired calibration
information and the probe serial number. This information is also included with all logged data.
Last DO Calibration Data
The last DO calibration data is stored automatically after a successful
calibration. To view the DO calibration data, press
GLP/CFM
key when
the instrument is in measurement mode.
The instrument will display:
The calibration standard and the calibration temperature:
0.0 % will be displayed if the instrument was calibrated at this point.
100.0 % calibration point, if instrument was calibrated in water‑
saturated air.
The altitude and salinity setting at the moment of calibration together
with the current reading.
The time of day that the calibration was performed together with the
current reading.
OPERA
TIONAL GUIDE