22
Logged data on the
edge
® DO
can be transferred from the meter to a PC by following these simple
directions. Suitable operating systems include Windows (XP minimum), OS X or Linux.
1. Connect
edge
® DO
to the PC using the supplied micro USB cable.
2. Turn on
edge
® DO
.
3. Press
SETUP/CLR
key and select “LOG ON EDGE”.
4. Press
CAL/MODIFY
key then use
keys to change to “EXPORT TO USB”.
5. Press
GLP/CFM
key and the USB/PC Tag is displayed.
6. Press
SETUP/CLR
key to exit.
The PC should detect the USB as a removable drive. Open the drive to view the stored files. Log
files are formatted as Comma separated values (*.CSV) and can be opened with any text editor or
spreadsheet program.
Note: Western Europe (ISO-88859-1) character set and English language are suggested
settings. Other files may be visible depending upon computer settings. All files stored will
appear in this folder.
Adjust Font or column width appropriately.
Interval logs are designated as DO Lots. ie. DOLOT001, DOLOT002, DOLOT003.
The Manual Lots are DOLOTMAN.
The Stability Lots are DOLOTSTAB. All stability logs, regardless of stability setting, are located in the
same stability file for that measurement.
Click on the desired log to view data.
Note:If “
°
C !” appeared in log data, the electrode/probe was used beyond it’s operation
specifications and the data is not considered reliable.If “
°
C !!” appeared in log data, the
temperature sensor within the probe or electrode is broken and the device should be replaced.
Logged data should not be considered reliable.