25
DO (Dissolved Oxygen) meter operation is configured using the SETUP key with a DO probe connected
to the meter. The parameter‑specific options will be seen inserted into the menu.
Parameter
Description
Choices
Default
Altitude (m)
Concentration
measurements of dissolved
oxygen change depending
on atmospheric pressure. A
convenient way to estimate
atmospheric pressure
effects is by the related
parameter of elevation
(m) above or below sea
level. Enter altitude in
meters closest to the actual
altitude to ensure the most
accurate calibration and
concentration measurement.
‑500, ‑400, ‑300, ‑200, ‑100,
0, 100, 200, 300, 400, 500,
600, 700, 800, 900, 1000,
1100, 1200, 1300, 1400,
1500, 1600, 1700, 1800,
1900, 2000, 2100, 2200,
2300, 2400, 2500, 2600,
2700, 2800, 2900, 3000,
3100, 3200, 3300, 3400,
3500, 3600, 3700, 3800,
3900, 4000 m
0
Salinity (g/L)
Dissolved oxygen solubility
decreases if water contains
salts. Selecting this factor
as to be close to your
known salt level, will
improve the accuracy of DO
concentration calibration
and measurement.
0, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10,
11, 12, 13, 14, 15, 16, 17
18, 19, 20, 21, 22, 23, 24,
25, 26, 27, 28, 29, 30, 31,
32, 33, 34, 35, 36, 37,38,
39, 40 g/L
0
DO Units
Select preferred
measurement units for DO
concentration.
mg/L or ppm
ppm
DISSOLVED OXYGEN METER CONFIGURATION
OPERA
TIONAL GUIDE
Summary of Contents for edge DO
Page 1: ...DO INSTRUCTION MANUAL ...