background image

6

7

PRINCIPLE OF OPERATION

Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation and
matter. When a light beam crosses a substance, some of the radiation may be absorbed by
atoms, molecules or crystal lattices.
If pure absorption occurs, the fraction of light absorbed depends both on the optical path length
through the matter and on the 

physical

-chemical characteristics of the substance according to the

Lambert-Beer Law:

-log

 

I

/

I

o

 = 

ε

λ

 

c d

or

A  

ε

λ

 

c d

Where:

-log

 

I

/

I

o

=

Absorbance (A)

I

o

=

intensity of incident light beam

I

=

intensity of light beam after absorption

ε

λ

=

molar extinction coefficient at wavelength

 

λ

c

=

molar concentration of the substance

d

=

optical path through the substance

Therefore, the concentration "c" can be calculated from the absorbance of the substance as the
other factors are known.
Photometric chemical analysis is based on the possibility to develop an absorbing compound
from a specific chemical reaction between sample and reagents. Given that the absorption of a
compound strictly depends on the wavelength of the incident light beam, a narrow spectral
bandwidth should be selected as well as a proper central wavelength to optimize measurements.
The optical system of Hanna's 

HI 96

 series colorimeters is based on special subminiature

tungsten lamps and narrow-band interference filters to guarantee both high performance and
reliable results.

HI 96 series block diagram (optical layout)

SPECIFICATIONS

PRECISION AND ACCURACY

Range

Free Cl

2

 0.00 to 5.00 mg/L

Total Cl

0.00 to 5.00 mg/L

Resolution

0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 mg/L above 3.50 mg/L

Accuracy

±0.03 mg/L ±3% of reading @ 25°C

Typical EMC Deviation

±0.01 mg/L

Light Source

Tungsten lamp

Light Detector

Silicon Photocell with narrow band interference filter @ 525nm.

Method

Adaptation of the USEPA method 330.5 and Standard Method 4500-Cl G.
The reaction between chlorine and the DPD reagent causes a pink tint
in the sample.

Environment

0 to 50°C (32 to 122°F); max 95% RH non-condensing

Battery Type

1 x 9 volt

Auto-Shut off

After 10' of non-use in 

measurement mode;

after 1 hour of non-use in 

calibration mode;

with last reading reminder.

Dimensions

192 x 104 x 69 mm (7.6 x 4.1 x 2.7")

Weight

360 g (12.7 oz.).

REQUIRED REAGENTS
Code                     Unit

        Description

Quantity/test

HI  93701-0            

Free Cl

2

        DPD Powder Reagent

1 packet

HI  93711-0      

      Total Cl

2

        DPD Powder Reagent

1 packet

Precision 

is how closely repeated measurements

agree with each other. Precision is usually
expressed as standard deviation (SD).

Accuracy

 is defined as the nearness of a test

result to the true value.
Although good precision suggests good accuracy,
precise results can be inaccurate. The figure
explains these definitions. In a laboratory using
a standard solution of 1.00 mg/L chlorine and a
representative lot of reagent, an operator obtained
with a single instrument a standard deviation of
0.02 mg/L.

Distributed by Applied Membranes, Inc. 

http://www.appliedmembranes.com

Distributed by Applied Membranes, Inc. 

http://www.appliedmembranes.com

Summary of Contents for 96711C

Page 1: ...1 Instruction Manual HI 96711C Free and Total Chlorine ISM ...

Page 2: ...GLP 23 BATTERY MANAGEMENT 24 BATTERY REPLACEMENT 24 ACCESSORIES 25 WARRANTY 26 Dear Customer Thank you for choosing a Hanna product This manual will provide you with the necessary information for the correct use of the instrument Please read it carefully before using the meter If you need additional technical information do not hesitate to e mail us at tech hannainst com All rights are reserved Re...

Page 3: ...ame position every time it is placed into the measurement cell It is designed to fit a cuvette with a larger neck making it easier to add both sample and reagents The cuvette is made from special optical glass to obtain best results The HI 96711 meter measures the free and total chlorine Cl2 content in water samples in the 0 00 to 5 00 mg L ppm range The method is an adaptation of the USEPA Method...

Page 4: ...sults HI 96 series block diagram optical layout SPECIFICATIONS PRECISION AND ACCURACY Range Free Cl2 0 00 to 5 00 mg L Total Cl2 0 00 to 5 00 mg L Resolution 0 01 mg L from 0 00 to 3 50 mg L 0 10 mg L above 3 50 mg L Accuracy 0 03 mg L 3 of reading 25 C Typical EMC Deviation 0 01 mg L Light Source Tungsten lamp Light Detector Silicon Photocell with narrow band interference filter 525nm Method Adap...

Page 5: ...tion of light It is recommended not to touch the cuvette walls with hands Furthermore in order to maintain the same conditions during the zeroing and the measuring phases it is necessary to close the cuvette to prevent any contamination INSTRUMENT DESCRIPTION KEYPAD DESCRIPTION ON OFF to turn the meter on and off ZERO CFM this is a bi functional key Just press to zero the meter prior to measuremen...

Page 6: ...ette are inverted Zero A zero reading was not taken Follow the instructions of the measurement procedure for zeroing the meter b on sample reading DISPLAY ELEMENTS DESCRIPTION 1 The measuring scheme lamp cuvette detector appears during different phases of zero or reading measurement 2 Error messages and warnings 3 The battery icon indicates the charge state of the battery 4 The hourglass appears w...

Page 7: ...tion is displayed normal operation of the instrument will be interrupted Change the battery and restart the meter Under range A blinking 0 00 indicates that the sample absorbs less light than the zero reference Check the procedure and make sure you use the same cuvette for reference zero and measurement Over Range A flashing value of the maximum concentration indicates an over range condition The ...

Page 8: ...cap must be tightened to the same degree Whenever the cuvette is placed into the measurement cell it must be dry outside and completely free of fingerprints oil or dirt Wipe it thoroughly with HI 731318 or a lint free cloth prior to insertion Shaking the cuvette can generate bubbles in the sample causing higher readings To obtain accurate measurements remove such bubbles by swirling or by gently t...

Page 9: ... turbidity or color the measurement takes place in two phases First the meter is zeroed using the unreacted sample After the reagents are added the reacted sample is measured Important note Free and Total Chlorine have to be measured separately with fresh unreacted samples if both values are requested MEASUREMENT PROCEDURE Turn the meter on by pressing ON OFF The display briefly shows all tags on ...

Page 10: ...ERFERENCES Bromine positive error Oxidized Manganese and Chromium positive error Chlorine dioxide positive error Ozone positive error Iodine positive error In case of water alkalinity greater than 250 mg L CaCO3 or acidity greater than 150 mg L CaCO3 the color of the sample may develop only partially or may rapidly fade To resolve this neutralize the sample with diluted HCl or NaOH In case of wate...

Page 11: ...rupt calibration procedure at any time by pressing CAL CHECK or ON OFF keys Warning Do not calibrate the instrument with standard solutions other than the HANNA CAL CHECKTM Standards otherwise erroneous results will be obtained When calibrating only the selected range is affected Turn the meter on by pressing ON OFF When the beeper sounds briefly and the LCD displays dashes the meter is ready To c...

Page 12: ...can be verified and the factory calibration can be restored LAST CALIBRATION DATE GLP Press RANGE GLP to edit the desired year 2009 2099 If the key is kept pressed the year number is automatically increased When the correct year has been set press ZERO CFM or READ TIMER to confirm Now the display will show the month blinking Press RANGE GLP to edit the desired month 01 12 If the key is kept presse...

Page 13: ... a new zero has to be performed One fresh battery lasts for around 750 measurements depending on the light level The remaining battery capacity is evaluated at the instrument startup and after each measurement The instrument displays a battery indicator with three levels as follows 3 lines for 100 capacity 2 lines for 66 capacity 1 line for 33 capacity Battery icon blinking if the capacity is unde...

Page 14: ... repair or replacement free of charge Damages due to accident misuse tampering or lack of prescribed maintenance are not covered If service is required contact your dealer If under warranty report the model number date of purchase serial number and the nature of the failure If the repair is not covered by the warranty you will be notified of the charges incurred If the instrument is to be returned...

Page 15: ...rintedinEUROPE ROMANIA 2450 Business Park Dr Vista CA 92081 760 727 3711 760 727 4427 www appliedmembranes com sales appliedmembranes com Industry Leader in RO Expertise and Membrane Applications since 1983 Distributed By ...

Reviews: