8 www.ihanil.com
3.
Installation
3.1
Packing inspection
3.2
Installation
▶
Check the delivery for completeness.
· Check packing conditions very carefully before unpacking.
· Contact Hanil Scientific immediately if damages found.
·
Check the delivery for completeness.
· You can get contact details on packing box or back of the manual.
▶
Check all parts for damage.
Installation on hard and flat ground.
• Centrifuge should be installed on hard and flat place.
• If the centrifuge is installed in an inclined place, the shaft may be bent due
to the weight of the rotor.
Good ventailation.
• For air circulation and safety, maintain a 30cm“clearance envelope”around
the centrifuge while the rotor is spinning.
Constant temperature/humidity
• Centrifuge equipped with the sensitive electronic software which is fragile
with humidity and temperature.
• Must avoid direct ray or heater and be put in the ambience of controlled
temp. and air.
Avoid the corrosive gas
• Install the centrifuge in a place where corrosive gas is not generated.
• Sulfur dioxide gas and chlorine gas may cause corrosion.
Leveling
• The shaft should be put exactly vertical on the hotizontaliy flat ground by
the leveling tool.