[email protected] / [email protected] 13
4.2
Setting
RPM
4.3
Setting run time
4.4
Starting/End the centrifugation
4.5
Pulse
function
4.
Operation
Rotate the 'RPM setting knob' clockwise or counterclockwise to set the rpm value.
▶
When using a A0.2-32 (PCR strip rotor) do not exceed 6,000 rpm.
Rotate 'Time setting knob' clockwise or counterclockwise to set the run time.
▶
0~15 min.
1. Check the all parameters are correct and the lid is closed.
2. Press the 'Start' key (Operation begins with a beep.)
▶
If the lid is not completely closed, it will not operate.
▶
When the set time is over, the device is decelerated and stops with beep sound.
▶
Press 'Stop' key to abort centrifugation.
▶
Set the 'Rpm setting knob' or 'Time setting knob' to '0' to abort centrifugation.
▶
When the lid is opened during operation, it decelerates and stops.
1. Set the 'Time setting knob' to '0'
2. Press the Start key.
3. The device decelerates and stops when the set rpm is reached.