4 www.ihanil.com
1.
General Considerations
1. The centrifuge should be installed on flat surface to maintain level.
2. Check the voltage to be used, before connecting the centrifuge to the power source.
4. Do not exceed the maximum rated speed of the rotor or buckets or accessories in use.
6. If the centrifuge is contaminated by toxic or radioactive samples or infectious blood samples, remove contaminants
completely
7. Substances that may generate volatile or explosive vapor can not be centrifuged.
8. The balancing work of samples shoud be done in advance before operation.
11. Only centrifuge with rotor and rotor lid firmly tightened.
12. Do not block vents.
Do not insert foreign objects into the holes of the device.
14. Repairs must be carried out by a technologist authorized by Hanil scientific Inc
.
15. According to IEC61010-2-020 maintain a 30cm “clearance envelope” around the centrifuge while the rotor is spinning.
16. Please contact the place of purchase or
Hanil Scientific Inc. ([email protected]/[email protected])
for
product
repairs.
1.2
Transport & Storage
1.1
Safety
Ambient temperature 5
℃
~ 35
℃
Maximum relative humidity 30% ~ 85%
Air pressure 500 ~ 1060 hpa
Follow precautions and all the safety requirements described on this user manual to prevent any
damages and failure of equipment and loss of lives.
Abnormal vibration is caused when the device is installed on an inclined surface.
3. Only use rotors, parts, and accessories provided by Hanil Scientific Inc.
Hanil
Scientific
Inc.
is
not
responsible
for
damages
of
the
device
and
accidents
caused
by
using
parts
and
accessories
not
recommended.
5. Make sure to prepare necessary safety measures before using samples that are toxic or radioactive samples or pathogenic
or samples or infectious blood.
and take needful actions such as ventilation or isolation of centrifuge.
9. Before operation, rotor and chamber should be dry.
10. Do not attempt to slow or stop the spinning rotor by hand.
13.
Storage
Transport
Ambient temperature -10
℃ ~ 40℃
Maximum relative humidity 10% ~ 90%
Air pressure 500 ~ 1060 hpa