After removing the RiseAtlas cover, check the following;
- All cables and connectors are correctly attached.
o If loose connector is found, attach correctly and reassemble the RiseAtlas covers.
- Disconnect the wiring of the battery and disassemble the batteries from the plastic brackets, see picture.
o Measure cross the black and red wire on the battery side, should be 24 – 25 VDC if fully charged.
- Attach the charger to the RiseAtlas unit and check charger voltage;
o Measure cross the black and red wire, should be 39 – 41 VDC when the charger is attached.
If there is correct charger voltage and low battery voltage – replace the battery with new battery, spare part no:
81000001.
Attach the new battery according to picture and connect terminal. Use cable tie to ensure that the battery cables don’t
touch with the gears. The old battery can be recycled at a recycling station as a NiMH-battery, environmental friendly.
Before reassembly the RiseAtlas covers, ensure that all connectors are correctly attached and that there are no
obstacles.
Set RiseAtlas on charge for at least 30 minutes and make a functional control of the lift unit according to chapter 2.3.2.
Notes of change should be presented in the lift unit log.
Please contact customer service if there are further issues regarding the battery change.
19
S E R V I C E M A N U A L
System
RoMedic
TM