Configuration Parameters
Note:
To configure target parameters, you must be logged in as an Administrator or Power User.
Click the CONFIGURE menu button
The Target Configuration Window
displays (see Figure 43). The following
fields are disabled (reserved for Hamar
Laser use only):
•
Device type
•
Serial
•
V KFactor
•
H KFactor
•
Mfg. Date
•
Cal. Station ID
•
Cal
•
Temp Compensation Factor
•
PV Cell Scale Factor
•
All Hardware Revision
•
Radio
AC Filter
Light Periodicity Synchronization
This setting is used by the incident light noise compensation filter to minimize the effects of background
light intensity fluctuations on the position sensor signal. The intensity of most light sources, such as
fluorescent lighting, fluctuates at the AC line frequency or twice the AC line frequency.
Set the Light Periodicity to the AC line frequency at the location where the target will be used as follows:
•
60/120Hz (United States)
•
50/100Hz or 60/120Hz (other countries)
Note:
DEBUG
is for Hamar Laser use only.
Avg. Samples
Internal Average Enabled. Position = Average of Samples
Vibration and air turbulence may cause target position readings to become noisy. You may have to
dampen the noise by averaging, particularly if the laser is far from the target. You may enable Target
Internal Averaging or Disable Internal Averaging if your application performs that function.
If Internal Averaging is enabled, the position reported by the target is the moving average of the last
n
position samples. You may select 2, 4, 8, 16, 32, 64 scans.
Figure 43 –
Target Configuration Window