34
c.
With buck-in offsets applied in the Bore9 software, steer the laser using the Pitch and Yaw knobs
(angular adjustment) on the L-706 until the
H
and
V
displays are zero.
d.
Repeat the process to confirm zero at both end bores. The laser is now “bucked in” (concentric) to
the centerline of the near and far reference bores.
Step 4: Record Data
a.
Move the target to the first bore you want to
measure for alignment and press
Record
. Continue
moving the target to each bore until all data is
taken.