16
Configuring the R-1307 or R-1307C for a Local Target
Before connecting and configuring a readout, make sure to mount the3 laser and install the target.
Note:
Shut off power to the readout before connecting or disconnecting a target from the local port.
1.
Connect the cabled target to the local port of the readout, as shown in Figure 14.
Refer to the R-1307 and R-1307C Control Panel drawings for the control panel buttons and indicator
functions applicable to your unit or to the R1307 manual for menu settings and further information.
2.
Set the Measurement Units
Press the MENU button until the upper display shows
Use the UP and DOWN arrow keys
to select either
for inches or
for millimeters.
3.
Set the Dampening Level
Press the MENU button until the upper display shows
. Use the UP and DOWN arrow keys to
set the number of averages. Four (4) is the default setting. Adjust this value as required to suit the
application.
4.
Set the Readout Function to Local Target
Press the MENU button until the upper display shows
. Use the UP and DOWN arrow keys
to select
.
5.
Select the PSD descriptor applicable to your target.
Press the MENU button until the upper display shows
. Use the UP and DOWN arrow
keys to select the Target Number. Refer to the R-1307 manual for more information regarding PSD
descriptors.
The following is applicable to the R-1307 only:
To make the unit visible to all other radio-enabled devices, you must set the Target Network ID and the
System ID for the readout.
1.
Set the Local Target Network ID
Press the MENU button until the
upper
display shows
with the current Target ID (
)
blinking. Use the UP and DOWN arrow keys to set the Target ID (the default is 01).
Press the MENU button again until the lower display shows
with the current Target ID
(
) blinking. Use the UP and DOWN arrow keys to set the Target ID to the same value as that of the
upper display’s Target ID (the default is 01).
The upper display will show the vertical (Y) axis position of the local target.
The lower display will show the Horizontal (X) axis position of the local target.
2.
Set the System ID (Radio Channel)
Press the MENU button until the upper display shows
, with the current System ID (
)
blinking. Use the UP and DOWN arrow keys to set the System ID (the default is 01).
Warning!
Targets are matched to specific calibration records in the R-1307 Readouts. For example, Target #1
must be connected to Calibration Record #1 in the R-1307 or the calibration is void. Up to nine
different target calibration records can be stored in each R-1307. When there are multiple calibration
records, the record ID must match the target ID, so if you have Target #1, you should select TGT=01
to select the matching calibration factors.
The R-1307B version of the readouts can only save one target calibration record, so each R-1307B
readout must be matched up the target the calibration factors were generated with.
Summary of Contents for L-705
Page 1: ......
Page 17: ...14 Model R 1307 Readout Control Panel...
Page 18: ...15 Model R 1307C Readout Control Panel...
Page 39: ...36 Bore9 Sample Report...
Page 44: ...41 Figure 25 Sample plot of readout data...
Page 47: ...44 Appendix D T 218 T 225L Dimensions Model 3...
Page 48: ...45 Appendix E T 225L Leg Dimensions Instructions...