10.5. Serial Interface Commands
The RS-485 serial interface is provided for user adjustment of laser operating power, in order to increase or
decrease light output. The EQ-77 is shipped with the laser power set to a factory default value of 100% of full
scale power. Laser power can be increased or decreased in increments of 0.5% of full scale. The following
describes the serial commands and their functions.
Commands consist of a single ASCII character, case-sensitive. This can be transmitted to the EQ-77 via a
terminal emulation program, or the user's control system. Response from the EQ-77 will be a string of ASCII
characters, format depending on the command issued.
The following describes the serial commands and their functions.
Command
Character
Function
Reply from EQ-77
U
Increases the present laser
power setpoint by 0.5% of full
scale
Power = XX.X%
D
Decreases the present laser
power setpoint by 0.5% of full
scale
Power = XX.X%
Q
Queries the value of the
present laser power setpoint
(in % of full scale)
Power = XX.X%
F
Resets laser power setpoint
to factory default value
Power = XX.X%
B
Saves present settings to
flash memory. This command
should be issued after the
settings are at their desired
values. If not, changes will be
lost if power to the EQ-77 is
interrupted.
Calibration data copied to FLASH
Memory.
?
Displays a help menu listing
the available commands
EQ77 firmware v02d (FRM-7117 R005)
build date 11/18/15
(c) Energetiq Technology, Inc.
operational commands:
U/D - raise/lower laser power by
0.5%
Q - Query present laser power
F - reset laser power to
Factory default
B - Burn present settings to
non-volatile flash
? - this help menu.