7
Located on the EQ-77 Controller front panel are LED system status indicators. The function of these
indicators is described below in Table
3.
LED Label
Meaning (when lit)
POWER OK
AC power is connected to the EQ-77 Power Supply Controller
LAMP ON
UV Light is on
LASER ON
Laser power is ON and laser light is being produced within the Lamp
CONTROLLER FAULT
One of the following has occurred in the Power Supply Controller:
External interlock open
Controller internal temperature too high
Laser power not reaching setpoint
Laser temperature fault
Internal power supply voltage low
LAMP MODULE FAULT
One of the following has occurred in the Lamp House module:
Control Cable not connected properly
Lamphouse internal temperature too high
Ignition Failure
Table 3:
Status Indicator LED Functions
3.2.2 Lamp House Assembly
The Lamp house assembly contains:
EQ-77 lamp
Monochromator
Optics
Interface connectors
Figure 3:
TLS-EQ-77H Lamp (rear view)