Specifications
Functions
Data transfer
Transfers the data stored in memory to the
computer.
Exposure time setting
1ms
〜
65535ms / 1ms step
External trigger input
BNC receptacle at rear panel.
Trigger mode setting
Internal trigger mode, External trigger
mode, External start mode
Trigger polarity setting
Positive, Negative
PULSE OUT setting
ON/OFF, Delay time, Pulse width
VIDEO OUT
BNC receptacle at rear panel.
SYNC. OUT
BNC receptacle at rear panel.
AMP GAIN setting
x1/2
、
x1
、
x2
、
x5
、
x10
、
x20
、
x50
、
x100
Peltier power supply setting
5V, 6V
Sensor head setting
CCD
、
InGaAs
、
N-MOS
Number of channel setting
Each maltichannel detector head is
available.
Sensor head mode setting
Operation mode, Load mode
The RS-232C interface is available.
Cooling Control
Cooling ON/OFF
Temperature data acquisition
Converts temperature signal to digital data,
and Transfers the data to the computer.